评价种植体初期稳定性的骨整合技术:尸体研究。

Evaluation of the Osseodensification Technique in Implant Primary Stability: Study on Cadavers.

出版信息

Int J Oral Maxillofac Implants. 2022 May-Jun;37(3):593-600. doi: 10.11607/jomi.9144.

Abstract

PURPOSE

The objective of this study was to compare the osseodensification drilling technique used with Densah implant drills (Versah) to standard drilling by evaluating their contribution to primary implant stability.

MATERIALS AND METHODS

A total of 21 mandibles dissected from cadavers were selected. Axiom drills (Anthogyr) were used for 29 standard drilling (SD) protocols in the control group. In the test group, the Densah profile drills were used to perform 29 osteotomies (OD). Implant placement was performed for 58 of these procedures. The insertion torque (IT) and implant stability quotient (ISQ) were recorded. High-resolution CBCT allowed for a postoperative analysis of the peri-drilling bone tissue. For 16 osteotomies without implant placement, Mann-Whitney U test was used for comparison of IT and ISQ values between groups. Spearman rank correlation coefficient was used between IT and ISQ values. The significance level was α = .05.

RESULTS

The IT values for OD and SD were, respectively, 34.9 Ncm ± 19.1 and 23.6 Ncm ± 9.8. The IT was significantly higher in the OD group compared to the SD group (P = .036). A moderate positive correlation (ρ = 0.527) between IT and ISQ was observed, as well as a significant increase (P = .026) in bone density.

CONCLUSION

There was a significant increase in IT and bone density following an osseodensification procedure compared to standard drilling.

摘要

目的

本研究旨在比较使用 Densah 种植钻(Versah)进行的骨密实化钻孔技术与标准钻孔技术,以评估其对初级种植体稳定性的贡献。

材料和方法

共选择了 21 个从尸体中解剖出的下颌骨。在对照组中,使用 Axiom 钻头(Anthogyr)进行 29 项标准钻孔(SD)方案。在试验组中,使用 Densah 轮廓钻头进行 29 个骨切开术(OD)。对其中 58 个程序进行了种植体放置。记录了植入扭矩(IT)和种植体稳定性指数(ISQ)。高分辨率 CBCT 允许对术后钻孔周围骨组织进行分析。对于 16 个没有植入物放置的骨切开术,使用 Mann-Whitney U 检验比较了两组之间的 IT 和 ISQ 值。使用 Spearman 秩相关系数比较了 IT 和 ISQ 值之间的相关性。显著性水平为α=0.05。

结果

OD 和 SD 的 IT 值分别为 34.9 Ncm±19.1 和 23.6 Ncm±9.8。OD 组的 IT 值明显高于 SD 组(P=0.036)。观察到 IT 和 ISQ 之间存在中度正相关(ρ=0.527),并且骨密度显著增加(P=0.026)。

结论

与标准钻孔相比,骨密实化程序后 IT 和骨密度显著增加。