Abstract

PURPOSE

Interstitial needles placement is a critical component of combined intracavitary/interstitial (IC/IS) brachytherapy (BT). To ensure precise placement of interstitial needles, we proposed a novel ultrasonic (US) probe calibration method to accurately register the US image in the magnetic resonance imaging (MRI) image and provide multimodal image guidance for needle placement.

METHODS

A wire-based calibration phantom combined with the stylus was developed for the calibration of US probe. The calibration phantom helps to quickly align the imaging plane of the US probe with the fiducial points to obtain US images of these points. The coordinates of fiducial points in US images were located automatically by feature extraction algorithms and were further corrected by the proposed correction method. Ingenious structures were designed on both sides of the calibration phantom to accurately obtain the coordinates of the fiducial points relative to the tracking device. Marker validation and pelvic phantom study were performed to evaluate the accuracy of the proposed calibration method.

RESULTS

In the marker validation, the US probe calibration method with corrected transformation achieves a registration accuracy of 0.694 ± 0.014 mm, and the uncorrected one is 0.746 ± 0.018 mm. In the pelvic phantom study, the needle tip difference was 1.096 ± 0.225 mm and trajectory difference was 1.416 ± 0.284 degrees.

CONCLUSION

The proposed US probe calibration method is helpful to achieve more accurate multimodality image guidance for needle placement.

摘要

目的

间质针放置是腔内/间质(IC/IS)近距离放射治疗(BT)的关键组成部分。为了确保间质针的精确放置,我们提出了一种新的超声(US)探头校准方法,以准确地将 US 图像与磁共振成像(MRI)图像进行配准,并为针的放置提供多模态图像引导。

方法

开发了一种基于线的校准体模与触笔相结合的方法,用于校准 US 探头。校准体模有助于快速将 US 探头的成像平面与基准点对齐,以获得这些点的 US 图像。通过特征提取算法自动定位基准点在 US 图像中的坐标,并通过提出的校正方法进一步校正。校准体模的两侧设计了巧妙的结构,以准确获得相对于跟踪装置的基准点坐标。进行了标记验证和骨盆体模研究,以评估所提出的校准方法的准确性。

结果

在标记验证中,经校正转换的 US 探头校准方法的配准精度为 0.694 ± 0.014mm,未经校正的配准精度为 0.746 ± 0.018mm。在骨盆体模研究中,针尖差异为 1.096 ± 0.225mm,轨迹差异为 1.416 ± 0.284 度。

结论

所提出的 US 探头校准方法有助于实现更精确的多模态图像引导针的放置。