Electrophoresis. 2022 Oct;43(18-19):1850-1858. doi: 10.1002/elps.202200134. Epub 2022 Jul 18.
Capillary electrophoresis with sodium dodecyl sulfate (CE-SDS) is a common analytical technique for investigating the purity and molecular size heterogeneity of monoclonal antibody (mAb) drugs. In reducing CE-SDS analysis of mAb-A, the light chain (LC) peak exhibited severe tailing, seriously affecting the purity analysis. The purposes of this investigation are to clarify the source of tailing and develop a more appropriate CE-SDS method to eliminate LC tailing. The degree of LC tailing was closely related to the mAb concentration, SDS concentration, and injection amount, and more hydrophobic detergents, such as sodium hexadecyl sulfate (SHS) and sodium tetradecyl sulfate (STS), could be used instead of SDS to obtain better peak shapes. The results also indicated that the tailing was caused by the binding problem associated with SDS, and SHS/STS could provide a more stable and uniform complexation for the LC. In summary, the method we developed successfully eliminated the LC tailing and provided a robust characterization of mAb-A in reducing CE-SDS analysis.
十二烷基硫酸钠毛细管电泳(CE-SDS)是一种常用于研究单克隆抗体(mAb)药物纯度和分子大小异质性的分析技术。在还原 CE-SDS 分析 mAb-A 时,轻链(LC)峰表现出严重的拖尾,严重影响了纯度分析。本研究旨在阐明拖尾的来源,并开发更合适的 CE-SDS 方法以消除 LC 拖尾。LC 拖尾的程度与 mAb 浓度、SDS 浓度和进样量密切相关,并且可以使用更疏水的洗涤剂,如十六烷基硫酸钠(SHS)和十四烷基硫酸钠(STS)代替 SDS 以获得更好的峰形。结果还表明,拖尾是由 SDS 结合问题引起的,SHS/STS 可以为 LC 提供更稳定和均匀的络合。总之,我们开发的方法成功消除了 LC 拖尾,并在还原 CE-SDS 分析中为 mAb-A 提供了稳健的表征。