Muhsen Sana, Roto Allaa, Al-Sabbagh Mohammed Qussay, Al-Huneidy Leen, AlRamahi Nura, Alhalaybeh Naja, Kotkot Duaa, Alkiswani Somaya, Al-Ani Abdallah, Abu-Yaghi Nakhleh
Special Surgery Department/Ophthalmology Division, School of Medicine, University of Jordan, Amman, Jordan.
Department of Neurology, University of Kansas Medical Center, Kansas City, KS, USA.
Eur J Ophthalmol. 2023 Jan;33(1):341-351. doi: 10.1177/11206721221113827. Epub 2022 Jul 15.
Th aim of this study is to explore the diagnostic accuracy of smartphone-based D-EYE ophthalmoscopy in the diagnosis of optic nerve head (ONH) abnormalities and screening for glaucoma.
This is a prospective clinical-based validation study performed on 90 patients recruited from a tertiary teaching hospital. Patients underwent dilated fundus examination by slit-lamp and smartphone-based D-EYE fundoscopy operated by two experienced ophthalmologists.
The diagnostic accuracy of smartphone-based D-EYE ophthalmoscopy of normal vertical cup-to-disc ratio (VCDR) was acceptable (sensitivity 85.8%-96.4%; specificity 51.4%-96.4%). The D-EYE would often underestimate VCDR values as mean VCDR was significantly lower among overall and glaucoma cohorts (all p-value <0.001) for D-EYE in comparison with slit-lamp. In terms of ONH abnormalities, the D-EYE adequately demonstrated high sensitivity in the identification of only margins-related abnormalities (sensitivity 92.5%-96.6%). Overall, the diagnostic accuracy and agreement between expert ophthalmologists using the D-EYE and the reference slit-lamp examination were significantly poor.
We demonstrated that the D-EYE is a highly specific tool for VCDR within the normal range but not in patients with glaucoma. Despite being an easy and portable tool to measure VCDR, it is limited in terms of diagnosing ONH and blood vessel abnormalities.
本研究旨在探讨基于智能手机的D-EYE检眼镜在诊断视神经乳头(ONH)异常及青光眼筛查中的诊断准确性。
这是一项基于临床的前瞻性验证研究,对从一家三级教学医院招募的90例患者进行。患者接受了由两名经验丰富的眼科医生操作的裂隙灯散瞳眼底检查和基于智能手机的D-EYE眼底镜检查。
基于智能手机的D-EYE检眼镜对正常垂直杯盘比(VCDR)的诊断准确性尚可(敏感性85.8%-96.4%;特异性51.4%-96.4%)。D-EYE常常会低估VCDR值,因为与裂隙灯相比,D-EYE检查的总体及青光眼队列中的平均VCDR显著更低(所有p值<0.001)。就ONH异常而言,D-EYE在仅识别边缘相关异常方面表现出较高的敏感性(敏感性92.5%-96.6%)。总体而言,使用D-EYE的专家眼科医生与参考裂隙灯检查之间的诊断准确性和一致性显著较差。
我们证明,D-EYE是一种对正常范围内的VCDR具有高度特异性的工具,但对青光眼患者并非如此。尽管它是一种测量VCDR的简便且便携的工具,但在诊断ONH和血管异常方面存在局限性。