BACKGROUND

Thyroglobulin (Tg) is an essential part for the management of patients with differentiated thyroid carcinoma (DTC) after thyroidectomy. Highly sensitive Tg assays are now established in clinical practice as they facilitate follow-up of DTC patients. In this study, we evaluated the recently launched highly sensitive Abbott Tg assay for Alinity and ARCHITECT.

METHODS

In this three-center study, Tg values of 447 routine patient samples, characterized for the presence of anti-Tg, were measured with the ARCHITECT Tg assay and compared with the Roche Elecsys TgII assay. In addition, a subset of 154 samples was compared also with the Beckman Tg Access assay and another subset (n = 122) with Abbott Tg on the Alinity i analyzer.

RESULTS

LoQ was verified to be less than or equal to 0.1 ng/ml confirming that the Tg assay on ARCHITECT and Alinity is highly sensitive. Correlation of ARCHITECT, Alinity, and Roche was excellent with a slope between 0.9 and 1.1 and a correlation coefficient >0.98. Correlation to Beckmann Tg was also very good, but the differences in absolute values were significant (slope: 1.477).

CONCLUSIONS

The Abbott Thyroglobulin assay, which is standardized to CRM-457, demonstrated a high correlation to the Roche and Beckman Tg assays, though good agreement of absolute values was only observed between Abbott and Roche. Strength of correlation and slope were not affected by the presence of anti-Tg indicating that all assays included in the study have a similar susceptibility to anti-Tg.

摘要

背景

甲状腺球蛋白(Tg)是甲状腺癌(DTC)患者甲状腺切除术后管理的重要组成部分。高敏 Tg 检测法现已在临床实践中确立,因为它有助于 DTC 患者的随访。在这项研究中,我们评估了新推出的 Abbott Alinity 和 ARCHITECT 高敏 Tg 检测法。

方法

在这项三中心研究中,对 447 例常规患者样本的 Tg 值进行了测定,这些样本具有抗 Tg 特征,使用 ARCHITECT Tg 检测法进行了检测,并与罗氏 Elecsys TgII 检测法进行了比较。此外,还比较了 154 例样本的贝克曼 Tg Access 检测法,以及另一组(n=122)Abbott Tg 检测法在 Alinity i 分析仪上的检测结果。

结果

验证了 LoQ 小于或等于 0.1ng/ml,证实 ARCHITECT 和 Alinity 上的 Tg 检测法具有高度的灵敏度。ARCHITECT、Alinity 和罗氏的相关性非常好,斜率在 0.9 到 1.1 之间,相关系数大于 0.98。与贝克曼 Tg 的相关性也非常好,但绝对值的差异非常显著(斜率:1.477)。

结论

与 CRM-457 标准化的 Abbott 甲状腺球蛋白检测法与罗氏和贝克曼 Tg 检测法具有高度的相关性,尽管仅在 Abbott 和罗氏之间观察到了绝对值的良好一致性。相关性和斜率的强度不受抗 Tg 的影响,表明研究中包含的所有检测法对抗 Tg 的敏感性相似。