J Orthop Surg Res. 2022 Jul 15;17(1):350. doi: 10.1186/s13018-022-03180-8.
To define the more stable knot tightening of a suture bridge when a single limb is preserved.
Five different sutures were tested: No. 2 Ethibond (Ethicon), Hi-Fi (ConMed Linvatec), Sutblue (SBM), SingleFlat (SBM), Hi-Fi Ribbon (ConMed Linvatec). A Surgeon's Knot was tied around a 30-mm circumference device, 6 times for each experiment. A single limb was kept to analyze failure modes of the knot. First step was to analyze which of pulling or sliding suture of the construct must be kept preventing failure of the knot. The cutting distance from the knot was evaluated at 1 mm and 4 mm with the suture loops pre-tensioned to 10 N and fixed to a second row after a 50 N tension load. The more stable construct was found: a single-pull load to 100 N and cyclic load (to 50 N for 30 cycles) experiments were conducted to evaluate the impact of cycling on knot loosening.
The more stable construct was obtained when the non-post limb was tensioned, and the post limb was cut at 4 mm (p < 0.01). Loop circumference increased after each experiment for all tested sutures, independently of the preserved limb and the cutting distance. Elongation was significant for all tested sutures in all groups. Knot failure mostly occurred by slippage, only with tapes.
A suture-bridge construct with the non-post limb preserved and the post limb cut at a 4 mm distance from the knot provides with the best security. Sutures are safer than tapes in suture bridge.
确定保留单肢时缝线桥更稳定的结扎。
测试了 5 种不同缝线:2 号爱惜邦(爱惜康)、Hi-Fi(康美林特维)、Sutblue(SBM)、SingleFlat(SBM)、Hi-Fi 丝带(康美林特维)。每个实验都在一个 30mm 周长的装置上打 6 个外科结。保留单肢以分析结的失效模式。首先分析必须保留缝线桥的哪个结构来防止结的失效,缝线桥的哪个结构被拉动或滑动。评估距离结 1mm 和 4mm 处的切割距离,缝线环预张力为 10N,并在 50N 张力负载后固定在第二排。发现更稳定的结构是:进行了单拉负载至 100N 和循环负载(30 个循环中至 50N)实验,以评估循环对结松动的影响。
当非后肢被拉紧,后肢在距离结 4mm 处被切割时,获得了更稳定的结构(p<0.01)。所有测试缝线的环周长在每次实验后都增加,与保留的肢和切割距离无关。所有测试缝线的伸长在所有组中均显著。结的失效主要是由于滑脱,仅发生在带材上。
保留非后肢并在后肢距离结 4mm 处切割的缝线桥结构提供了最佳安全性。缝线在缝线桥中比带材更安全。