Suppr超能文献

髋关节表面置换植入物患者血液金属离子水平的长期变化:10 年随访后患者监测的意义。

Long-term changes in blood metal ion levels in patients with hip resurfacing implants: implications for patient surveillance after 10 years follow-up.

机构信息

Coxa Hospital for Joint Replacement, Tampere, Finland and Faculty of Medicine and Health Technologies, University of Tampere, Tampere, Finland.

The Royal Orthopaedic Hospital, Birmingham, UK.

出版信息

Hip Int. 2023 Sep;33(5):864-871. doi: 10.1177/11207000221104396. Epub 2022 Jul 17.

Abstract

BACKGROUND

Studies investigating changes in blood metal ion levels during the second decade of the implant lifetime in MoM hip resurfacing patients are scarce.

METHODS

Patients implanted with either Birmingham Hip Resurfacing (BHR) or Articular Surface Replacement (ASR) hip resurfacings with >10 years follow-up and repeated blood metal ion measurements were identified at 2 large specialist European arthroplasty centres. After excluding patients with initial metal ion levels >7 ppb, the proportion of patients with an increase in blood metal ion levels above previously validated implant-specific thresholds (cobalt 2.15 ppb for unilateral implants, cobalt or chromium 5.5 ppb for bilateral) was assessed.

RESULTS

We included 2743 blood metal ion measurements from 457 BHR patients (555 hips) and 216 ASR patients (263 hips). Of patients with initial metal ion levels below implant specific thresholds, increases in cobalt or chromium level, respectively, to above these thresholds during the second decade were seen as follows: unilateral BHR (cobalt = 15.6%), unilateral ASR (cobalt = 13.8%), bilateral BHR (cobalt = 8.2%, chromium = 11.8%), bilateral ASR (cobalt = 8.5%, chromium = 4.3%). Measurement-to-measurement changes exceeding +2.15 ppb or +5.5 ppb were, however, uncommon during the second decade. Subgroup results with small diameter (<50 mm) implants were similar.

CONCLUSIONS

We recommend less frequent blood metal ion measurements are needed (every 3 to 5 years) for hip resurfacing patients if initial values were below 7ppb.

摘要

背景

在 MoM 髋关节表面置换患者中,研究植入后第二个十年期间血液金属离子水平变化的研究很少。

方法

在 2 家大型欧洲关节置换中心,确定了随访时间>10 年且重复进行血液金属离子测量的 Birmingham Hip Resurfacing(BHR)或 Articular Surface Replacement(ASR)髋关节表面置换患者。在排除初始金属离子水平>7 ppb 的患者后,评估了血液金属离子水平升高超过先前验证的特定植入物阈值(单侧植入物钴 2.15 ppb,双侧钴或铬 5.5 ppb)的患者比例。

结果

我们纳入了 457 名 BHR 患者(555 髋)和 216 名 ASR 患者(263 髋)的 2743 次血液金属离子测量值。在初始金属离子水平低于植入物特定阈值的患者中,在第二个十年期间,钴或铬水平分别升高到这些阈值以上,结果如下:单侧 BHR(钴=15.6%),单侧 ASR(钴=13.8%),双侧 BHR(钴=8.2%,铬=11.8%),双侧 ASR(钴=8.5%,铬=4.3%)。然而,在第二个十年期间,测量值到测量值的变化超过+2.15 ppb 或+5.5 ppb 的情况并不常见。小直径(<50mm)植入物的亚组结果相似。

结论

如果初始值低于 7ppb,我们建议髋关节表面置换患者需要更频繁地测量血液金属离子(每 3 至 5 年一次)。