Detecting sleep apnea in adults with Down syndrome using WatchPAT: A feasibility study.

机构信息

Department of Practice-oriented Scientific Research (PWO), Alliade, Heerenveen, The Netherlands; Department of Health Sciences, Applied Health Research, University of Groningen, University Medical Center Groningen, Groningen, The Netherlands.

Department of Practice-oriented Scientific Research (PWO), Alliade, Heerenveen, The Netherlands.

出版信息

Res Dev Disabil. 2022 Oct;129:104302. doi: 10.1016/j.ridd.2022.104302. Epub 2022 Jul 16.

Abstract

BACKGROUND

In daily practice, sleep apnea is underdiagnosed in people with Down syndrome. The WatchPAT can detect sleep apnea in a less invasive way.

AIM

This study aimed to evaluate the feasibility of the WatchPAT to detect sleep apnea in individuals with Down syndrome.

METHODS AND PROCEDURES

Thirty-one participants with Down syndrome (aged 18+) were included. Sleep apnea was detected with the WatchPAT and compared to results of the STOP-Bang Questionnaire (current practice). Experiences of participants, caregivers and clinicians were studied using a combination of quantitative and qualitative methods.

OUTCOMES AND RESULTS

Among the 68% of participants who accepted the WatchPAT, sleep apnea was detected in 95% of participants. Younger participants and participants with mild/moderate intellectual disabilities were more likely to accept the device. STOP-Bang did not detect most cases of sleep apnea. For the degree of sleep apnea, interrater reliability was substantial (k = 0.71) to almost perfect (k = 0.91). Considering experiences, caregivers and clinicians were predominantly positive about the WatchPAT.

CONCLUSIONS

Our study showed that the WatchPAT is a promising device to detect sleep apnea in people with Down syndrome. Compared to polysomnography, detection with this device is less invasive and less burdensome for people with Down syndrome. Furthermore, the WatchPAT is a relatively accessible solution to implement in care institutions.

摘要

背景

在日常实践中,唐氏综合征患者的睡眠呼吸暂停症常常被漏诊。WatchPAT 可以以一种侵入性更小的方式来检测睡眠呼吸暂停症。

目的

本研究旨在评估 WatchPAT 检测唐氏综合征患者睡眠呼吸暂停症的可行性。

方法和程序

共纳入 31 名唐氏综合征患者(年龄 18 岁以上)。使用 WatchPAT 检测睡眠呼吸暂停症,并与 STOP-Bang 问卷(当前实践)的结果进行比较。使用定量和定性相结合的方法研究了参与者、照顾者和临床医生的经验。

结果和结论

在接受 WatchPAT 检测的 68%的参与者中,有 95%的参与者检测出睡眠呼吸暂停症。年轻的参与者和智力障碍程度较轻/中度的参与者更有可能接受该设备。STOP-Bang 并未检测到大多数睡眠呼吸暂停症。对于睡眠呼吸暂停症的严重程度,组内相关系数为 0.71(高度可靠)至 0.91(几乎完美)。考虑到经验,照顾者和临床医生对 WatchPAT 的评价主要是积极的。

我们的研究表明,WatchPAT 是一种有前途的检测唐氏综合征患者睡眠呼吸暂停症的设备。与多导睡眠图相比,使用该设备对唐氏综合征患者的侵入性更小,负担更小。此外,WatchPAT 是一种相对容易在护理机构中实施的解决方案。