Noise is a common problem in wearable electrocardiogram (ECG) monitoring systems because the presence of noise can corrupt the ECG waveform causing inaccurate signal interpretation. By comparison with electromagnetic interference and its minimization, the reduction of motion artifact is more difficult and challenging because its time-frequency characteristics are unpredictable. Based on the characteristics of motion artifacts, this work uses adaptive filtering, a specially designed ECG device, and an Impedance Pneumography (IP) data acquisition system to combat motion artifacts. The newly designed ECG-IP acquisition system maximizes signal correlation by measuring both ECG and IP signals simultaneously using the same pair of electrodes. Signal comparison investigations between ECG and IP signals under five different body motions were carried out, and the Pearson Correlation Coefficient |r| was higher than 0.6 in all cases, indicating a good correlation. To optimize the performance of adaptive motion artifact reduction, the IP signal was filtered to a 5 Hz low-pass filter and then fed into a Recursive Least Squares (RLS) adaptive filter as a reference input signal. The performance of the proposed motion artifact reduction method was evaluated subjectively and objectively, and the results proved that the method could suppress the motion artifacts and achieve minimal distortion to the denoised ECG signal.
噪声是可穿戴心电图 (ECG) 监测系统中的一个常见问题,因为噪声的存在会使 ECG 波形失真,导致信号解释不准确。与电磁干扰及其最小化相比,运动伪影的减少更加困难和具有挑战性,因为其时频特性是不可预测的。基于运动伪影的特点,本工作采用自适应滤波、专门设计的 ECG 设备和阻抗式呼吸描记术 (IP) 数据采集系统来对抗运动伪影。新设计的 ECG-IP 采集系统通过使用同一对电极同时测量 ECG 和 IP 信号,最大限度地提高了信号相关性。对五种不同身体运动下的 ECG 和 IP 信号进行了信号比较研究,在所有情况下,Pearson 相关系数 |r| 均高于 0.6,表明相关性良好。为了优化自适应运动伪影减少的性能,将 IP 信号过滤到 5 Hz 的低通滤波器,并将其作为参考输入信号输入到递归最小二乘 (RLS) 自适应滤波器中。主观和客观评估了所提出的运动伪影减少方法的性能,结果证明该方法可以抑制运动伪影,并对去噪后的 ECG 信号实现最小的失真。