Shear wave elastography is a valuable tool for diagnosing and grading carpal tunnel syndrome.

机构信息

Department of Radiology, Hospital Sírio-Libanês, R. Adma Jafet, 91, São Paulo, 01308-050, Brazil.

出版信息

Skeletal Radiol. 2023 Jan;52(1):67-72. doi: 10.1007/s00256-022-04143-0. Epub 2022 Aug 3.

Abstract

OBJECTIVE

To quantify the shear velocity and stiffness of the median nerve (MN) with shear wave elastography (SWE) at the carpal tunnel entrance and determine whether SWE is useful for diagnosing and staging carpal tunnel syndrome (CTS).

MATERIALS AND METHODS

The study included 58 patients (79 wrists) with clinical and electroneuromyographic diagnoses of CTS and 55 healthy controls (63 wrists). MN shear velocity and stiffness were measured by SWE on the axial plane in both groups. The differences between CTS patients and controls and between different grades of CTS based on electrodiagnostic tests were studied using Student's t test and ANOVA with ROC analysis.

RESULTS

The mean MN shear velocity and stiffness were significantly greater in CTS patients (2.5 ± 0.37 m/s and 19.4 ± 5.8 kPa) than in controls (1.91 ± 0.24 m/s and 11.1 ± 3.0 kPa) (p < 0.001) and greater in the severe CTS group (2.69 ± 0.39 m/s and 22.4 ± 7.1 kPa) than in the mild CTS group (2.37 ± 0.35 m/s and 17.3 ± 4,8 kPa). The cutoff value for the shear velocity was 2.13 m/s, with 86% and 82% sensitivity and specificity, respectively, and the cutoff value for stiffness was 13.6 kPa, with 87% and 82% sensitivity and specificity.

CONCLUSION

MN shear velocity and stiffness are significantly higher in CTS patients. SWE can be used to diagnose CTS and distinguish between patients with mild and severe disease.

摘要

目的

使用剪切波弹性成像(SWE)定量测量腕管入口处正中神经(MN)的剪切波速度和杨氏模量,并确定 SWE 是否有助于诊断和分期腕管综合征(CTS)。

材料与方法

本研究纳入了 58 例(79 只腕)经临床和电生理诊断为 CTS 的患者和 55 例健康对照者(63 只腕)。在两组中,均在轴向平面上使用 SWE 测量 MN 的剪切波速度和杨氏模量。使用 Student's t 检验和 ANOVA 分析 ROC 曲线,研究 CTS 患者与对照组之间以及根据电诊断测试不同 CTS 分级之间的差异。

结果

CTS 患者的 MN 剪切波速度和杨氏模量明显高于对照组(分别为 2.5±0.37 m/s 和 19.4±5.8 kPa;1.91±0.24 m/s 和 11.1±3.0 kPa)(p<0.001),重度 CTS 组的 MN 剪切波速度和杨氏模量也明显高于轻度 CTS 组(分别为 2.69±0.39 m/s 和 22.4±7.1 kPa;2.37±0.35 m/s 和 17.3±4.8 kPa)。剪切波速度的截断值为 2.13 m/s,其敏感性和特异性分别为 86%和 82%,杨氏模量的截断值为 13.6 kPa,其敏感性和特异性分别为 87%和 82%。

结论

CTS 患者的 MN 剪切波速度和杨氏模量明显升高。SWE 可用于诊断 CTS,并区分轻度和重度疾病患者。