Ind Health. 2023 Jul 29;61(4):260-268. doi: 10.2486/indhealth.2022-0046. Epub 2022 Aug 8.
This study (1) evaluated the perceptual and objective physical quality of digital radiographic chest images processed for different purposes (routine hospital use, lung cancer screening, and pneumoconiosis screening), and (2) quantified objectively the quality of chest images visually graded by the Japan National Federation of Industrial Health Organization (ZENEIREN). Four observers rated the images using a visual grading score (VGS) according to ZENEIREN's quality criteria. Signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR) were measured. Between groups, differences were assessed using ANOVA (followed by Bonferroni multiple comparisons) or unpaired t-test. The Pearson's correlation coefficients were calculated for the correlation between perceptual quality and objective physical image quality. The image quality perceived by the observers and the SNR measurements were highest for the images generated using parameters recommended for lung cancer screening. The images processed for pneumoconiosis screening were rated poorest by the observers and showed the lowest objective physical quality measurements. The chest images rated high quality by ZENEIREN generally showed a higher objective physical image quality. The SNR correlated well with VGS, but CNR did not. Highly significant differences between the processing parameters indicate that image processing strongly influences the perceptual quality of digital radiographic chest images.
本研究(1)评估了为不同目的(常规医院使用、肺癌筛查和尘肺病筛查)处理的数字射线照相胸部图像的感知和客观物理质量,(2)使用日本全国产业卫生联合会(ZENEIREN)的视觉分级法(ZENEIREN)对胸部图像的质量进行客观量化。四位观察者根据 ZENEIREN 的质量标准使用视觉分级评分(VGS)对图像进行评分。测量信噪比(SNR)和对比噪声比(CNR)。使用方差分析(随后进行 Bonferroni 多重比较)或未配对 t 检验评估组间差异。计算了感知质量与客观物理图像质量之间的相关性的 Pearson 相关系数。使用推荐用于肺癌筛查的参数生成的图像,观察者感知的图像质量和 SNR 测量值最高。观察者评价最差的是为尘肺病筛查处理的图像,其客观物理质量测量值最低。ZENEIREN 评价为高质量的胸部图像通常具有更高的客观物理图像质量。SNR 与 VGS 相关性良好,但 CNR 没有。处理参数之间存在显著差异,表明图像处理强烈影响数字射线照相胸部图像的感知质量。