评估锥形束 CT 中定位片功能的剂量学和中心化。

Evaluation of the dosimetry and centralization of scout-view function in CBCT.

机构信息

Division of Oral Radiology, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil.

Division of Cell Biology, Faculdade São Leopoldo Mandic, Campinas, São Paulo, Brazil.

出版信息

Braz Dent J. 2022 Jul-Aug;33(4):31-39. doi: 10.1590/0103-6440202204926.

Abstract

This study evaluated the centralization of the region of interest (ROI) in acquisition of the CBCT images, when the freely positionable scout-view (SV) function is applied. Additionally, the dosimetry of the acquired images was assessed in the SV function alone as well as in complete tomographic image in two different fields of view (FOV) (50x50 and 78x150mm). A three-location device was created to accommodate the dosimeters and the specimens, in the right, middle and left location during image acquisition. For dose assessment, thermoluminescent dosimeters were irradiated within the FOV and analyzed in a portable reader. For ROI evaluation, three specimens of gutta-percha stick were placed on the same device and the CT scans were acquired (CBCT OP 300 Maxio device, 90kV, 13mA, 85 µm voxel size, FOV of 50X50mm), with and without the SV, in three positions (3-9, 1-7 and 5-11 o'clock), simulating different regions of the mouth. Two image evaluations were performed, an objective and subjective. There was a slight percentage increase (1.36% to 1.40%) of the radiation dose with the use of SV. The distances were significantly greater in the images acquired without SV (p < 0.05). Every image obtained with SV was classified as being at the FOV's center. In conclusion, the results demonstrated that SVs function is effective to centralize the ROI in the FOV, increasing the scan precision and avoiding repetitions due to positioning errors.

摘要

本研究评估了在应用可自由定位的 scout-view(SV)功能时,感兴趣区域(ROI)在 CBCT 图像采集时的集中程度。此外,还单独评估了 SV 功能以及在两个不同视野(FOV)(50x50 和 78x150mm)中获取的完整断层图像中的剂量。创建了一个三位置装置,以在图像采集期间在右侧、中间和左侧位置容纳剂量计和标本。为了进行剂量评估,将热释光剂量计置于 FOV 内并在便携式读取器中进行分析。为了评估 ROI,将三根牙胶棒样本放置在同一装置上,并进行 CT 扫描(CBCT OP 300 Maxio 设备,90kV,13mA,85µm 体素大小,50x50mm 的 FOV),分别在三个位置(3-9、1-7 和 5-11 点)进行扫描,模拟口腔的不同区域。进行了两次图像评估,客观评估和主观评估。使用 SV 后,辐射剂量略有增加(1.36%至 1.40%)。在没有 SV 采集的图像中,距离明显更大(p<0.05)。使用 SV 获得的每张图像都被分类为处于 FOV 的中心。总之,结果表明 SV 功能可有效将 ROI 集中在 FOV 内,提高扫描精度并避免因定位错误而重复。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/7eb2/9645184/30c2f1409f85/1806-4760-bdj-33-04-31-gf1.jpg

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验