Barrett R, Hyde S A, Hark W B
J Med Eng Technol. 1987 Mar-Apr;11(2):68-73. doi: 10.3109/03091908709062798.
A force platform was built that was shown to meet the design and operating specifications for clinical use in the application of techniques of stabilography to investigate disturbances of equilibrium in young boys with Duchenne muscular dystrophy. The force platform was tested for reliability and repeatability and results were at an acceptable level for clinical application. Further, the design of the force platform and the associated software for system operation and computation of results were usable by non-technical staff. A feasibility study reporting the effect of the application of orthoses to a group of eight boys with Duchenne muscular dystrophy, who had lost the ability to walk independently, is represented.
搭建了一个测力平台,经证明该平台符合临床使用的设计和操作规范,可用于应用稳定图技术研究患有杜氏肌营养不良症的年轻男孩的平衡障碍。对该测力平台进行了可靠性和重复性测试,结果达到了临床应用可接受的水平。此外,非技术人员也可以使用该测力平台的设计以及用于系统操作和结果计算的相关软件。本文展示了一项可行性研究,该研究报告了对一组八名失去独立行走能力的杜氏肌营养不良症男孩应用矫形器的效果。