Rev Fac Cien Med Univ Nac Cordoba. 2022 Sep 16;79(3):298-300. doi: 10.31053/1853.0605.v79.n3.32651.

Abstract

BACKGROUND

The phenomenon of deformation of devices for closure of intracardiac defects in catheterization has been reported, but not of devices for closure of patent ductus arteriosus.

OBJECTIVE

to report a case of deformation of a relatively new type of device for closure of patent ductus arteriosus.

METHODS

report an adult with hypertensive patent ductus arteriosus and a positive balloon occlusion test will be presented, in whom the occlusion was attempted with an Occlutech®-PDA device.

RESULTS

Upon delivery, the device took the shape of a “horn” instead of a “champagne cork” (usual appearance), so it was extracted prior to it is release and the procedure was carried out with another device.

CONCLUSION

To our knowledge, no deformation during the delivery transcatheter of the Occlutech®-PDA device has been published. Adverse events related to new technologies, as in our case, should be reported as soon as possible, since it can help manufacturers to improve their quality and provide greater safety to patients.

摘要

背景

心内缺损封堵装置在导管插入术中出现变形的现象已有报道,但动脉导管未闭封堵装置的变形情况尚未见报道。

目的

报告一例相对新型的动脉导管未闭封堵装置发生变形的病例。

方法

报告一名患有高血压性动脉导管未闭且球囊封堵试验阳性的成年人,尝试使用Occlutech®-PDA装置进行封堵。

结果

输送时,该装置呈“喇叭”状而非“香槟瓶塞”状(正常外观),因此在释放前将其取出,改用另一装置进行手术。

结论

据我们所知,尚未有关于Occlutech®-PDA装置经导管输送过程中发生变形的报道。与新技术相关的不良事件,如我们所报道的病例,应尽快报告,因为这有助于制造商提高产品质量,为患者提供更高的安全性。