Xu Huiting, Zhao Yanjun, Yuan Jiaqi, Li Wei, Ni Jianming
Department of Radiology, Wuxi No. 2 Peolpe's Hospital, Affiliated Wuxi Clinical College of Nangtong University, Wuxi, Jiangsu 214042, China.
Can Assoc Radiol J. 2023 May;74(2):455-461. doi: 10.1177/08465371221133482. Epub 2022 Oct 27.
To evaluate a novel laser angle selection system (LASS) for improving the efficiency of a computed tomography (CT)-guided percutaneous transthoracic needle biopsy (PTNB). Thirty-eight patients referred for CT-guided PTNB were randomly separated into a LASS-assisted puncture group (18 patients) or conventional freehand control group (20 patients). The puncture time, number of control CT scans, and patients' radiation dose were compared for each group. The lesion size, target-to-pleural distance, planned puncture depth, and angle of the two groups were not significantly different. LASS-assisted PTNB significantly reduced the number of control scans (1.7 ± 0.8 vs 3.5 ± 1.5, < .001) and the mean operation time (12.0 ± 4.3 min vs 28.8 ± 13.3 min, < .001) compared with the conventional method. The corresponding room time (27.1 ± 6.6 min vs 44.1 ± 14.4 min, < .001) and total radiation dose (7.9 ± 1.0 mSv vs 10.1 ± 1.7 mSv, < .001) of each procedure also decreased significantly. Fifty-six percent (10/18) of the operations hit the target on the first needle pass when using LASS compared with 10% (2/20) using the conventional method. Compared with a conventional method, this novel laser angle simulator improves puncture efficiency with fewer needle readjustments and reduces patient radiation dose.
评估一种新型激光角度选择系统(LASS)以提高计算机断层扫描(CT)引导下经皮经胸针吸活检(PTNB)的效率。38例因CT引导下PTNB就诊的患者被随机分为LASS辅助穿刺组(18例患者)或传统徒手对照组(20例患者)。比较每组的穿刺时间、对照CT扫描次数和患者的辐射剂量。两组的病变大小、靶点至胸膜距离、计划穿刺深度和角度无显著差异。与传统方法相比,LASS辅助PTNB显著减少了对照扫描次数(1.7±0.8 vs 3.5±1.5,<0.001)和平均手术时间(12.0±4.3分钟 vs 28.8±13.3分钟,<0.001)。每个操作的相应操作间时间(27.1±6.6分钟 vs 44.1±14.4分钟,<0.001)和总辐射剂量(7.9±1.0 mSv vs 10.1±1.7 mSv,<0.001)也显著降低。使用LASS时,56%(10/18)的操作在首次进针时命中靶点,而使用传统方法时为10%(2/20)。与传统方法相比,这种新型激光角度模拟器通过更少的针调整提高了穿刺效率,并降低了患者的辐射剂量。