Microbiol Spectr. 2022 Dec 21;10(6):e0312522. doi: 10.1128/spectrum.03125-22. Epub 2022 Oct 31.
The cobas 5800 System ("cobas 5800") is a new low- to mid-throughput PCR-based nucleic acid testing system which performs both qualitative and quantitative testing, including viral load (VL) determination. cobas 5800 shares numerous design elements and technical characteristics with the existing cobas 6800/8800 Systems. We compared HBV, HCV, and HIV-1 VL results from cobas 5800 in three different laboratories to those from the same specimens tested on a cobas 6800 system. We also assessed cobas 5800 assay reproducibility by repetitive testing of specimens with VL close to values used as thresholds for patient management or classification. The correlation between VL measurements generated using cobas 5800 versus 6800 was extremely high, with r correlation coefficients between 0.990 and 0.999 for the three targets at the different sites. The slope of the Deming regression line ranged from 0.994 (HBV, site 3) to 1.025 (HIV-1, site 1). The standard deviation values ranged from 0.04 to 0.19 log IU/mL for HBV, 0.06 to 0.33 log IU/mL for HCV, and 0.05 to 0.34 log copies/mL for HIV-1. In general, variability was higher at lower VL. Between 98.6% and 100% of results fell within the allowable total difference zone that defines expected variability on the existing 6800/8800 system. This multisite comparison study demonstrates equivalent performance of the new cobas 5800 system compared with cobas 6800. This establishes cobas 5800 as a new option for low- to mid-throughout laboratories seeking to optimize efficiency of their viral molecular testing. These are the first published data that demonstrate equivalent performance of the new cobas 5800 system compared with cobas 6800. This fulfills an unmet need for low- to mid-throughout laboratories seeking to optimize efficiency of their viral molecular testing.
cobas 5800 系统(“cobas 5800”)是一种新的中低通量基于 PCR 的核酸检测系统,可进行定性和定量检测,包括病毒载量(VL)测定。 cobas 5800 与现有的 cobas 6800/8800 系统有许多设计元素和技术特点。我们比较了三个不同实验室的 cobas 5800 中 HBV、HCV 和 HIV-1 VL 结果与在 cobas 6800 系统上测试相同标本的结果。我们还通过对接近用于患者管理或分类的阈值的 VL 接近值的标本进行重复测试,评估了 cobas 5800 检测的可重复性。使用 cobas 5800 与 6800 生成的 VL 测量之间的相关性非常高,三个靶标在不同地点的 r 相关系数在 0.990 到 0.999 之间。Deming 回归线的斜率范围从 0.994(HBV,站点 3)到 1.025(HIV-1,站点 1)。标准偏差值范围为 HBV 为 0.04 至 0.19 log IU/mL,HCV 为 0.06 至 0.33 log IU/mL,HIV-1 为 0.05 至 0.34 log 拷贝/mL。一般来说,在较低的 VL 时,变异性更高。98.6%至 100%的结果都在现有 6800/8800 系统定义的允许总差异范围内,这表明结果是可以接受的。这项多站点比较研究表明,新的 cobas 5800 系统与 cobas 6800 系统的性能相当。这为寻求优化其病毒分子检测效率的中低通量实验室提供了一个新的选择。这些是首次发表的数据,表明新的 cobas 5800 系统与 cobas 6800 系统的性能相当。这满足了中低通量实验室优化其病毒分子检测效率的未满足需求。
Expert Rev Mol Diagn. 2019-6-19
Rev Chilena Infectol. 2021-6
World J Virol. 2023-9-25
Micromachines (Basel). 2024-1-17
Diagnostics (Basel). 2021-2-12
J Clin Virol. 2020-8
J Clin Virol. 2020-8
Expert Rev Mol Diagn. 2019-9-17