Mattiello Christopher J, Stickle Douglas F
Jefferson University Hospital, Philadelphia, PA, United States.
Jefferson University Hospital, Philadelphia, PA, United States.
Clin Chim Acta. 2023 Jan 1;538:175-180. doi: 10.1016/j.cca.2022.11.020. Epub 2022 Nov 22.
As an adjunct to verification of performance characteristics of a qualitative serum hCG lateral flow immunoassay (LFI), we performed image analysis to characterize the dose vs response curve (visibility of the test line), as a means of understanding the transition from negative to positive as a function of increasing [hCG].
Using serum samples of known [hCG], device images were obtained using a scanner at the prescribed reading time (5 min). Image analysis (using Python and R) was used to obtain the integral (S) of the test-line color as a function of [hCG].
Data for S as a function of [hCG] were well characterized by a simple hyperbola: S = Smax [hCG]/([hCG] + K), where K = 202 mIU/ml (r = 0.997). Replicates of S at K had CV of 7.3 %. By eye, uncertainty of test results among users occurred only below the assay's stated sensitivity of 10 mIU/ml, in region of S < 3 % of Smax, and signal:noise ratio < 3.
By image analysis, the dose vs response (Test line integral) for this qualitative serum hCG LFI was a simple hyperbola. Characterization of the dose vs response curve was useful in verification of the assay's performance characteristics.
作为定性血清人绒毛膜促性腺激素侧向流动免疫分析(LFI)性能特征验证的辅助手段,我们进行了图像分析以表征剂量与反应曲线(测试线的可见性),作为理解随着[hCG]增加从阴性到阳性转变的一种方法。
使用已知[hCG]的血清样本,在规定的读取时间(5分钟)使用扫描仪获取设备图像。使用图像分析(使用Python和R)来获取测试线颜色积分(S)作为[hCG]的函数。
S作为[hCG]函数的数据可以用一个简单的双曲线很好地表征:S = Smax[hCG]/([hCG] + K),其中K = 202 mIU/ml(r = 0.997)。K处S的重复测量值的CV为7.3%。肉眼观察,用户之间测试结果的不确定性仅发生在该分析规定的10 mIU/ml灵敏度以下,即S < Smax的3%且信噪比< 3的区域。
通过图像分析,这种定性血清hCG LFI的剂量与反应(测试线积分)为简单双曲线。剂量与反应曲线的表征有助于验证该分析的性能特征。