Suppr超能文献

采用体积排阻微采样装置提取全血后,通过高效液相色谱-二极管阵列检测法定量检测非那韦。

HPLC-DAD quantification of favipiravir in whole blood after extraction from volumetric absorptive microsampling devices.

机构信息

Laboratory of Bioavailability/Bioequivalence, Faculty of Pharmacy, Universitas Indonesia, Depok, Indonesia.

Laboratory of Bioavailability/Bioequivalence, Faculty of Pharmacy, Universitas Indonesia, Depok, Indonesia; Laboratory of Pharmaceutical Chemistry, Faculty of Military Pharmacy, Republic of Indonesia Defense University, Bogor, Indonesia.

出版信息

J Chromatogr B Analyt Technol Biomed Life Sci. 2023 Jan 15;1215:123547. doi: 10.1016/j.jchromb.2022.123547. Epub 2022 Dec 12.

Abstract

Favipiravir is a prodrug of T-1105 made by modifying the pyrazine group as a COVID-19 therapy. During the pandemic, a safe and comfortable biosampling technique is needed for the subject or patient. Volumetric Absorptive Microsampling (VAMS) is a biosampling technique with a small blood volume and minimum hematocrit effect. The aims of this study were to develop and validate an analytical method for quantifying favipiravir extracted from VAMS using High Performance Liquid Chromatography - Photodiode Array with remdesivir as an internal standard. Analysis of favipiravir was performed using a C column (Waters, Sunfire™ 5 µm; 250 × 4.6 mm), with injection volume of 50 µL, flow rate of 0.8 mL/min, column temperature 30 ℃, and wavelength 300 nm. The separation was conducted under gradient elution with mobile phase consists of acetonitrile-0.2 % formic acid-20 mM sodium dihydrogen phosphate pH 3.5 and run time 12 min. Sample preparation was carried out using a protein precipitation method with 500 µL of methanol as precipitating agent. Samples were mixed on vortex for 30 s, sonicated for 15 min, and centrifuged at 10,000 rpm for 10 min. Lower Limit of Quantification (LLOQ) obtained was 0.5 µg/mL and the calibration curve ranged from 0.5 to 160 µg/mL. Sensitivity, linearity, selectivity, carry-over, accuracy, precision, recovery, and stability were validated by the guideline from Food and Drug Administration 2018. The method developed has successfully met the full validation requirements by FDA 2018 with the LLOQ obtained was 0.5 µg /mL.

摘要

法匹拉韦是一种由吡嗪基团修饰而成的 T-1105 的前药,作为 COVID-19 治疗药物。在大流行期间,需要为受试者或患者提供安全舒适的生物样本采集技术。体积吸收微采样(VAMS)是一种使用小体积血液和最小血细胞比容影响的生物样本采集技术。本研究的目的是开发和验证一种从 VAMS 中提取法匹拉韦并使用高效液相色谱-光电二极管阵列进行定量分析的方法,以内标瑞德西韦进行分析。法匹拉韦的分析使用 C 柱(沃特世,Sunfire™ 5 µm;250×4.6 mm)进行,进样体积为 50 µL,流速为 0.8 mL/min,柱温为 30℃,波长为 300nm。分离在梯度洗脱下进行,流动相由乙腈-0.2%甲酸-20 mM 磷酸二氢钠 pH 3.5 组成,运行时间为 12 分钟。样品制备采用 500 µL 甲醇沉淀法进行,沉淀剂。样品在涡旋混合 30 s,超声处理 15 min,然后以 10,000 rpm 离心 10 min。获得的定量下限(LLOQ)为 0.5 µg/mL,校准曲线范围为 0.5 至 160 µg/mL。灵敏度、线性、选择性、拖尾、准确性、精密度、回收率和稳定性均按照 2018 年美国食品和药物管理局的指南进行验证。所开发的方法已成功满足美国食品和药物管理局 2018 年的全面验证要求,LLOQ 为 0.5 µg/mL。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/2545/9743781/0edd522ff5bf/gr1_lrg.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验