γ相机线扩展函数测量中的实际考量

Practical considerations in gamma camera line spread function measurement.

作者信息

Tyson R K, Amtey S R

出版信息

Med Phys. 1978 Nov-Dec;5(6):480-4. doi: 10.1118/1.594457.

Abstract

In recent years the modulation transfer function (MTF) has played an important role in the quantitation of imaging performance of gamma cameras. The most common method of MTF determination requires line spread function (LSF) measurement. This paper reviews methods used for LSF measurements with special consideration given to the practical aspects of LSF measurement and MTF calculation. An analysis of errors in LSF measurements is made and means to reduce or to avoid these errors are discussed. Recommendations regarding practical considerations for LSF measurement and MTF calculation are presented in tabular form for convenience.

摘要

近年来,调制传递函数(MTF)在γ相机成像性能定量分析中发挥了重要作用。MTF测定的最常用方法需要测量线扩展函数(LSF)。本文回顾了用于LSF测量的方法,并特别考虑了LSF测量和MTF计算的实际问题。对LSF测量中的误差进行了分析,并讨论了减少或避免这些误差的方法。为方便起见,以表格形式给出了关于LSF测量和MTF计算实际考虑因素的建议。