Suppr超能文献

使用自行采集和医护人员采集的样本,通过视觉判读以及BD Veritor™ Plus分析仪检测抗原的2019冠状病毒病检测方法的准确性。

Accuracy of an antigen-detecting SARS-CoV-2 assay with self- and provider-collected specimens interpreted visually and with BD Veritor™ Plus analyzer.

作者信息

Cooper Lauren, Eckert Karen, Mann Joseph, Montalvo Rossmeri, Castillo Luis, Fellows Dwan, Young Stephen

机构信息

Becton Dickinson and Company, Integrated Diagnostic Solutions, 7 Loveton Circle, Sparks, MD, USA.

CTMD Research, 2328 S. Congress Ave, Suite 1-C, Palm Springs, FL, USA.

出版信息

J Clin Virol Plus. 2023 Feb;3(1):100140. doi: 10.1016/j.jcvp.2023.100140. Epub 2023 Jan 18.

Abstract

BACKGROUND

SARS-CoV-2 rapid antigen tests (RATs) are in high demand for reducing the spread of SARS-CoV-2. Reduced involvement from health care professionals (HCPs) for collection and interpretation could significantly foster the wide-spread implementation of RATs, but data evaluating RATs, when used by lay people, is limited.

OBJECTIVE

To valuate agreement between BD Veritor test results for self- and HCP-collected specimens, and visually- and analyzer-interpreted results.

METHODS

Individuals with onset of COVID-19 symptoms within five days of enrollment had three nasal swabs collected; one self-collected and the other two HCP-collected. One HCP-collected swab was stored for future testing while the order of the other two (self and HCP) was randomized before testing. with the BD Veritor System for Rapid Detection of SARS-CoV-2. Results were first assessed visually, followed by interpretation with the analyzer.

RESULTS

When self-collection was compared to HCP collection for SARS-CoV-2 detection, interpretation by analyzer resulted in positive percent agreement (PPA) of 94.7% (95% CI 82.7, 98.5) and negative percent agreement (NPA) of 99.0% (95% CI 97.5, 99.6). When visual interpretation was compared to analyzer-read results, collection by HCPs had a PPA of 97.4% (95% CI 86.5, 99.5) and NPA of 99.8% (95% CI 98.6, 100.0) while self-collection resulted in PPA of 94.9% (95% CI 83.1, 98.6) and NPA of 99.8% (95% CI 98.6, 100).

CONCLUSIONS

Similar PPA and NPA were observed for self- and HCP-collected specimens as well as visually- and analyzer-interpreted tests. The equivalence in performance supports the use of expanded collection and testing methods.

摘要

背景

严重急性呼吸综合征冠状病毒2(SARS-CoV-2)快速抗原检测(RAT)对于减少SARS-CoV-2传播的需求很高。医疗保健专业人员(HCP)在样本采集和结果解读方面参与度的降低可以显著促进RAT的广泛应用,但关于非专业人员使用RAT的数据有限。

目的

评估BD Veritor检测自我采集和HCP采集样本的结果之间,以及视觉解读和分析仪解读结果之间的一致性。

方法

在入组后五天内出现COVID-19症状的个体采集三份鼻拭子样本;一份自我采集,另外两份由HCP采集。一份HCP采集的拭子样本保存以备后续检测,另外两份(自我采集和HCP采集)在检测前随机排序,使用BD Veritor系统进行SARS-CoV-2快速检测。结果首先进行视觉评估,然后用分析仪进行解读。

结果

将自我采集与HCP采集用于SARS-CoV-2检测进行比较时,分析仪解读的阳性百分比一致性(PPA)为94.7%(95%CI 82.7,98.5),阴性百分比一致性(NPA)为99.0%(95%CI 97.5,99.6)。将视觉解读与分析仪读取结果进行比较时,HCP采集的PPA为97.4%(95%CI 86.5,99.5),NPA为99.8%(95%CI 98.6,100.0),而自我采集的PPA为94.9%(95%CI 83.1,98.6),NPA为99.8%(95%CI 98.6,100)。

结论

自我采集和HCP采集的样本以及视觉解读和分析仪解读的检测结果观察到了相似的PPA和NPA。性能上的等效性支持使用扩展的采集和检测方法。