Strakas Dimitrios, Dionysopoulos Dimitrios, Tolidis Kosmas, Meister Jörg
Department of Operative Dentistry, Faculty of Dentistry, School of Health Sciences, Aristotle University of Thessaloniki, Thessaloniki, Greece.
Center of Dento-Maxilo-Facial Medicine, Center of Applied Medical Laser Research and Biomedical Optics (AMLaReBO), Bonn University, Bonn, Germany.
Lasers Surg Med. 2023 Mar;55(3):294-304. doi: 10.1002/lsm.23639. Epub 2023 Feb 5.
To investigate quantitatively the cutting efficiency and the thermal effects in the surrounding soft tissues of incisions that are induced by a 940 nm-diode laser with different power settings.
Fifty-four gingival samples were prepared from the lower jaws of freshly slaughtered German-land race pigs and were randomly divided into 9 groups (n = 6) according to the adjusted output power (1, 1.5, 2, 2.5, 3, 3.5, 4, 5 and 6 W). Five incisions were implemented for each sample using a diode laser (940 nm) in continuous wave with an initiated tip resulting in 30 incisions for each experimental group utilizing a three-dimensional computer-controlled micropositioner. The samples were prepared for histometric evaluation using a transmitted light microscope. The cutting depth and width and the thermal damage were recorded for each sample and the efficiency factor γ was calculated.
The highest cutting efficiency (γ = 0.81 ± 0.03) exhibited the group with 5 W output power (p < 0.05), while the lowest (γ = 0.45 ± 0.11) showed the 1-W group (p < 0.05). Over 3.5 W there was a rapid increase in the size of thermal damage of the incisions, especially for 6 W, which presented the largest.
The most effective power parameters of diode laser (940 nm) for soft tissue surgery were from 3 to 5 W. The outcomes of the current study may help to establish clinical protocols for the use of diode lasers (940 nm) in soft tissue surgery in contact mode assisting dental professionals to achieve optimal clinical results and avoid complications.
定量研究不同功率设置的940nm二极管激光所产生的切割效率以及切口周围软组织的热效应。
从刚屠宰的德国长白猪下颌骨制备54个牙龈样本,根据调整后的输出功率(1、1.5、2、2.5、3、3.5、4、5和6W)随机分为9组(n = 6)。使用二极管激光(940nm)以连续波模式对每个样本进行5次切口,通过三维计算机控制微定位器,每个实验组共产生30个切口。使用透射光显微镜对样本进行组织测量评估。记录每个样本的切割深度、宽度和热损伤情况,并计算效率因子γ。
输出功率为5W的组切割效率最高(γ = 0.81±0.03)(p < 0.05),而1W组最低(γ = 0.45±0.11)(p < 0.05)。超过3.5W时,切口热损伤大小迅速增加,尤其是6W组,热损伤最大。
二极管激光(940nm)用于软组织手术的最有效功率参数为3至5W。本研究结果可能有助于制定二极管激光(940nm)在接触模式下用于软组织手术的临床方案,协助牙科专业人员获得最佳临床效果并避免并发症。