Fujiwara M, Ishida Y, Nimura N, Toyama A, Kinoshita T
Analytical Instrument Division, Shimadzu Corporation, Kyoto, Japan.
Anal Biochem. 1987 Oct;166(1):72-8. doi: 10.1016/0003-2697(87)90547-1.
A high-performance liquid chromatographic system for the determination of amino acid and imino acid was developed using a two-step reaction with sodium hypochlorite and o-phthalaldehyde/N-acetyl-L-cysteine (OPTA/AcCys) reagent. This reagent improved the sensitivity in the analysis of proline presumably because the fluorophore is more stable to hypochlorite which has been used for the oxidative cleavage of the imino linkage. The use of OPTA/AcCys facilitated the detection of imino acids at the same concentration level as that of amino acids. The detection limit for all the amino and imino acids was a few picomoles. This detection system, together with cation-exchange chromatographic separation, was applied to the determination of amino and imino acids in biological samples.
开发了一种用于测定氨基酸和亚氨基酸的高效液相色谱系统,该系统采用与次氯酸钠和邻苯二甲醛/N-乙酰-L-半胱氨酸(OPTA/AcCys)试剂的两步反应。该试剂提高了脯氨酸分析的灵敏度,可能是因为荧光团对用于亚氨基键氧化裂解的次氯酸盐更稳定。OPTA/AcCys的使用有助于在与氨基酸相同的浓度水平下检测亚氨基酸。所有氨基酸和亚氨基酸的检测限为几皮摩尔。该检测系统与阳离子交换色谱分离相结合,用于测定生物样品中的氨基酸和亚氨基酸。