Air entrapment as a cause of early inappropriate shocks after subcutaneous defibrillator implant: A case series.

作者信息

Iavarone Michele, Ammendola Ernesto, Rago Anna, Russo Vincenzo

机构信息

Cardiology Unit, Department of Translational Medical Sciences, University of Campania "Luigi Vanvitelli", Monaldi Hospital, Naples, Italy.

Cardiology Unit, Department of Translational Medical Sciences, University of Campania "Luigi Vanvitelli", Monaldi Hospital, Naples, Italy.

出版信息

Indian Pacing Electrophysiol J. 2023 May-Jun;23(3):84-87. doi: 10.1016/j.ipej.2023.02.001. Epub 2023 Feb 4.

Abstract

Inappropriate shock (IAS) is the most frequent device-related complication among patients with subcutaneous implantable cardioverter-defibrillator (S-ICD). Air entrapment (AE) has been described as an underdiagnosed cause of early postimplant IAS. We report 6 consecutive cases of early IAS after S-ICD implant, in whom the electrogram analysis (EGM) and/or chest radiography (CXR) were consistent with AE.

摘要

不适当电击(IAS)是皮下植入式心律转复除颤器(S-ICD)患者中最常见的与设备相关的并发症。空气栓塞(AE)已被描述为植入后早期IAS的一种诊断不足的原因。我们报告了6例连续的S-ICD植入后早期IAS病例,其心电图分析(EGM)和/或胸部X线摄影(CXR)结果与AE一致。