J Pediatr Urol. 2023 Jun;19(3):248.e1-248.e6. doi: 10.1016/j.jpurol.2023.01.009. Epub 2023 Jan 19.
To compare the efficacy, side effects, and cost-effectiveness between a single-use digital flexible ureteroscope and a reusable flexible ureteroscope in the treatment of paediatric renal stones.
This analytic, case-control, monocentric study included all patients undergoing flexible ureterosopies for stone treatment. Between April 2016 and February 2019, a reusable (Flex-X®, Karl Storz) flexible ureteroscope was used (control group), whereas a single-use (Uscope®, PUSEN Medical©) flexible ureteroscope was used in all procedures from March 2019 to April 2021. Clinical and procedural outcomes, operative times, complication rates, hospital stay, and costs per procedure were evaluated.
Forty-three cases using a reusable flexible ureteroscope and thirty-nine using a single-use flexible ureteroscope were included in the study. Demographic patient characteristics, stone burden, location and composition, preoperative presence of a double-J stent, procedural outcomes, mean length of postoperative hospital stay, and complications (4.6% versus 5%, p = 0.81) were comparable between the two groups. Median operative duration for stone removal was 93 min (20-170) with reusable versus 81 min (55-107) with the single-use scope (p = 0.18). Scope failure occurred four times with the reusable scope and in no case with the single-use. The total cost per procedure associated with the use of single-use scopes (798 Euros) was lower than a reusable scope (1483.23 Euros).
Single-use flexible ureteroscopes were created to bypass the problems incurred when reusable scopes were damaged and therefore not available for use in surgical procedures. Single-use flexible ureteroscopes are always immediately available and ready to be used, even in urgent cases, as they typically do not require maintenance or sterilization. Compared with their reusable counterparts, single-use flexible ureteroscopes have similar digital performance (270°), image quality and we found no difference in the success and complication rates. Cost analysis of a reusable flexible ureteroscope must consider the purchase price, maintenance and repair costs, and decontamination costs (including handling, detergent, bacterial culture, transportation, and storage costs). In contrast, only purchase price is included in cost analysis for single-use flexible ureteroscopes. Our study suggests that single-use flexible ureteroscopes may be associated with lower costs per procedure than their reusable counterparts.
Single-use flexible ureteroscopes are an interesting alternative to their reusable counterparts, particularly in terms of material resource management. Cost analyses conducted using a low volume of cases representative of a paediatric urology division favour the use of single-use ureteroscopes.
比较单次使用的数字输尿管软镜和可重复使用的输尿管软镜在治疗小儿肾结石中的疗效、副作用和成本效益。
本分析性、病例对照、单中心研究纳入了所有接受输尿管软镜治疗结石的患者。2016 年 4 月至 2019 年 2 月期间使用可重复使用的(Flex-X®,Karl Storz)输尿管软镜(对照组),而 2019 年 3 月至 2021 年 4 月期间所有手术均使用一次性使用的(Uscope®,PUSEN Medical©)输尿管软镜。评估临床和手术结果、手术时间、并发症发生率、住院时间和每例手术的成本。
本研究纳入了 43 例使用可重复使用输尿管软镜和 39 例使用一次性使用输尿管软镜的病例。两组患者的人口统计学特征、结石负荷、位置和成分、术前双 J 支架的存在、手术结果、术后平均住院时间和并发症(4.6%对 5%,p=0.81)相似。可重复使用输尿管软镜组结石清除的中位手术时间为 93 分钟(20-170 分钟),而一次性使用输尿管软镜组为 81 分钟(55-107 分钟)(p=0.18)。可重复使用输尿管软镜组有 4 次出现镜体故障,而一次性使用输尿管软镜组则没有。单次使用输尿管软镜(798 欧元)的每次手术相关总成本低于可重复使用输尿管软镜(1483.23 欧元)。
一次性使用输尿管软镜的出现是为了避免在可重复使用的输尿管软镜损坏且无法用于手术时出现的问题。一次性使用输尿管软镜始终可立即使用,即使在紧急情况下也无需等待,因为它们通常无需维护或消毒。与可重复使用的输尿管软镜相比,一次性使用输尿管软镜具有相似的数字性能(270°)、图像质量,我们发现手术成功率和并发症发生率没有差异。可重复使用输尿管软镜的成本分析必须考虑购买价格、维护和维修成本以及去污成本(包括处理、清洁剂、细菌培养、运输和储存成本)。相比之下,一次性使用输尿管软镜的成本分析仅包括购买价格。我们的研究表明,单次使用输尿管软镜的每次手术相关成本可能低于可重复使用输尿管软镜。
一次性使用输尿管软镜是可重复使用输尿管软镜的一个有趣替代品,特别是在材料资源管理方面。使用具有代表性的小儿泌尿科部门的低病例量进行的成本分析支持使用一次性输尿管软镜。