University of Milan-Bicocca, A.S.S.T. Monza, San Gerardo Hospital, Monza (MB), Italy.
A.S.S.T. Monza, San Gerardo Hospital, Monza (MB), Italy.
J Vasc Access. 2024 Jul;25(4):1187-1193. doi: 10.1177/11297298231153517. Epub 2023 Feb 10.
Correct tip positioning is a critical aspect in central vascular access devices insertion. The verification of positioning at the cavo-atrial junction is usually performed by intracavitary electrocardiography. Recently, echocardiographic techniques were proposed, including the direct visualization of the catheter or the visualization of a saline/air bolus (i.e. "bubble test"). As for the latter, a push-to-bubbles delay time below 2 s was proposed to indicate a correct positioning of the catheter tip. The aim of this study was to measure the variations of the push-to-bubbles time at increasing distance from the cavo-atrial junction, to verify if a cut-off of 1-2 s correspond to a well-positioned catheter.
We performed a prospective study including patients with clinical indication of positioning a peripherally inserted central catheter. The catheter tip was positioned at the cavo-atrial junction (P0) via intracavitary electrocardiography, and the push-to-bubbles delay time was measured. The catheter was then retracted 5 cm (P1) and 10 cm (P2), and the test was repeated at this positioning. Push-to-bubbles time measurements were performed off-line by analyzing an audio/video recording which included the echography screen and the voice signal of the operator.
Forty-nine patients were included. The average push-to-bubble time when the catheter tip was in the reference position was 0.41 ± 0.21 s. Retraction of the PICC catheter of 5 and 10 cm determined a significant increase of the push-to-bubbles time: mean time difference was +0.34 (95% IC 0.25-0.43, < 0.001) s between P0 and P1 (5 cm distance), and +0.77 (95% IC 0.62-0.92, < 0.001) s between P0 and P2 (10 cm distance). When the catheter was at the reference position (i.e. cavo-atrial junction) only 2.1% of bubbles delay times were above 1 s.
The push-to-bubbles time is very low when the catheter tip is at the cavo-atrial junction. This delay increases progressively with increasing distance from the target. Push-to-bubbles delay time above 1 s might indicate a catheter not close to the cavo-atrial junction.
正确的尖端定位是中央血管通路装置插入的关键方面。通常通过心腔内心电图来验证在腔静脉-心房交界处的定位。最近,提出了超声心动图技术,包括导管的直接可视化或盐水/空气团块的可视化(即“气泡测试”)。对于后者,提出了推注至气泡的延迟时间低于 2 秒,以指示导管尖端的正确定位。本研究的目的是测量从腔静脉-心房交界处增加距离时推注至气泡的时间的变化,以验证 1-2 秒的截止值是否对应于定位良好的导管。
我们进行了一项前瞻性研究,纳入了有中央外周插入的导管定位临床指征的患者。通过心腔内心电图将导管尖端定位在腔静脉-心房交界处(P0),并测量推注至气泡的延迟时间。然后将导管回缩 5 厘米(P1)和 10 厘米(P2),并在此位置重复测试。通过分析包括超声心动图屏幕和操作员语音信号的音频/视频记录,离线进行推注至气泡时间测量。
纳入了 49 名患者。当导管尖端处于参考位置时,推注至气泡的平均时间为 0.41±0.21 秒。PICC 导管回缩 5 和 10 厘米会导致推注至气泡时间的显著增加:P0 与 P1(5 厘米距离)之间的平均时间差为+0.34(95%置信区间 0.25-0.43,<0.001)秒,P0 与 P2(10 厘米距离)之间的平均时间差为+0.77(95%置信区间 0.62-0.92,<0.001)秒。当导管处于参考位置(即腔静脉-心房交界处)时,只有 2.1%的气泡延迟时间超过 1 秒。
当导管尖端位于腔静脉-心房交界处时,推注至气泡的时间非常短。随着与目标距离的增加,这种延迟会逐渐增加。推注至气泡的延迟时间超过 1 秒可能表明导管未靠近腔静脉-心房交界处。