The Patellostabilometer: A New Device for Quantification of Mediolateral Patella Displacement.

机构信息

Center of Orthopaedics and Traumatology, Brandenburg Medical School Theodor Fontane, University Hospital Brandenburg/Havel, 14770 Brandenburg a.d.H., Germany.

Technical Faculty, Brandenburg University of Applied Sciences, 14770 Brandenburg a.d.H., Germany.

出版信息

Sensors (Basel). 2023 Jan 22;23(3):1274. doi: 10.3390/s23031274.

Abstract

Mediolateral patella displacement is of interest for diagnostics and clinically relevant research questions. Apart from manual testing, no standardized method is currently available. Proper quantification of patella mobility is necessary to better understand pathologies at the patellofemoral joint. Patella mobility was assessed in 25 healthy individuals using a Patellostabilometer, a new prototype instrument for quantification of the mediolateral patella displacement. The participants underwent measurements of the mediolateral displacement three times using the Patellostabilometer. A maximal force of 10 N was applied for patella movement. Additionally, leg length and circumference of the knee, upper- and lower-leg were measured. Lateral patella displacement of 18.27 ± 3.76 mm (range 15.85-20.64 mm, interquartile range (IQR) of 4.79) was measured. The medial patella displacement showed 24.47 ± 6.59 mm (range 19.29-29.76 mm, IQR of 10.47). The test-retest measurement error was 2.32 ± 1.76 mm (IQR of 2.38 mm), with five outliers. There was greater test-retest variability between the measurements of the medial displacement compared to the lateral one. The test-retest variability reached 7% of the patella displacement. Other parameters provided no significant correlations. Based on the natural patellofemoral mobility, a precise and clinically relevant quantification of patella mobility is allowed.

摘要

髌股外侧位移对于诊断和临床相关研究问题很重要。除了手动测试外,目前尚无标准化方法。为了更好地了解髌股关节的病理学,需要对髌股的活动度进行适当的量化。使用髌股稳定仪评估了 25 名健康个体的髌股外侧位移,髌股稳定仪是一种用于量化髌股外侧位移的新型原型仪器。参与者使用髌股稳定仪进行了 3 次髌股外侧位移测量。施加 10 N 的最大力以使髌骨移动。此外,还测量了腿长和膝关节的周长、上下腿的周长。测量到的外侧髌股位移为 18.27 ± 3.76 mm(范围 15.85-20.64 mm,四分位距(IQR)为 4.79)。内侧髌股位移为 24.47 ± 6.59 mm(范围 19.29-29.76 mm,IQR 为 10.47)。测试-再测试测量误差为 2.32 ± 1.76 mm(IQR 为 2.38 mm),有 5 个离群值。内侧位移的测量比外侧位移的测试-再测试变异性更大。测试-再测试的变异性达到髌股位移的 7%。其他参数没有显著相关性。基于自然髌股的活动性,可以对髌股的活动度进行精确且具有临床相关性的量化。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/6a7f/9920919/84531003af50/sensors-23-01274-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验