The effect of different computed tomography (CT) scanning methods of the shoulder on image quality is uncertain.
To compare the effect of different methods of CT scanning of the right shoulder on image quality and radiation dose.
A total of 30 adults were divided into five groups. Group A received scans centered on the body's long axis, a scout direction of 0° + 90°, and automatic tube current modulation (ATCM). The other four groups (B, C, D, E) received isocenter scans centered on the shoulder with different scout directions (B and C: 0° + 90°, D: 0°, E: 0° + 270°) and tube currents (B: 420 mA; C, D, E: ATCM). The volume CT dose index (CTDI), dose-length product (DLP), image objective noise, signal-to-noise ratio (SNR), and contrast-to-noise ratio (CNR) were compared. Three subjective measures were also compared (noise, stripe artifacts, diagnostic confidence).
The five groups differed significantly in all subjective and objective indexes. The CTDI and DLP decreased in the order of groups C, A, B, E, and D; the differences between groups A and B were not significant ( > 0.05). Groups B, C, and E had better SNR and CNR than groups A and D ( < 0.01). Subjective evaluations indicated group D was worse than groups B, C, and E ( < 0.05).
In the ATCM system that uses the last scout view, CT of the shoulder should use isocenter scanning with the lateral scout view when the tube is away from the long axis of the body as the last execution direction.
不同的肩部 CT 扫描方法对图像质量的影响尚不确定。
比较不同的右肩关节 CT 扫描方法对图像质量和辐射剂量的影响。
共纳入 30 名成年人,分为 5 组。A 组采用基于身体长轴的扫描,扫描前的探测方向为 0°+90°,并使用自动管电流调制(ATCM)。其余 4 组(B、C、D、E)采用基于肩关节的等中心扫描,不同的探测方向(B 和 C:0°+90°,D:0°,E:0°+270°)和管电流(B:420mA;C、D、E:ATCM)。比较容积 CT 剂量指数(CTDI)、剂量长度乘积(DLP)、图像客观噪声、信噪比(SNR)和对比噪声比(CNR)。还比较了 3 种主观测量指标(噪声、条纹伪影、诊断信心)。
5 组在所有主观和客观指标上均有显著差异。CTDI 和 DLP 按组 C、A、B、E、D 的顺序递减;组 A 和 B 之间的差异无统计学意义(>0.05)。组 B、C 和 E 的 SNR 和 CNR 均优于组 A 和 D(<0.01)。主观评估显示,组 D 劣于组 B、C 和 E(<0.05)。
在使用最后一次探测视图的 ATCM 系统中,当管电流远离身体长轴作为最后执行方向时,肩关节 CT 应采用基于侧面探测视图的等中心扫描。