Five-year evaluation of a low-cost quality assurance protocol for clinical ultrasound transducers.

作者信息

Hyldgaard Niels, Bolander Malvang Lise, Brix Lau

机构信息

Department of Radiology, Diagnostic Centre, University Research Clinic for Innovative Patient Pathways, Silkeborg Regional Hospital, Silkeborg, Denmark.

Department of Procurement and Biomedical Engineering, Region Midt, Aarhus N, Denmark.

出版信息

Ultrasound. 2023 Feb;31(1):71-78. doi: 10.1177/1742271X221091721. Epub 2022 May 23.

Abstract

OBJECTIVE

To implement, execute and evaluate a quality assurance test protocol for clinical ultrasound transducers that is software-based, user-independent, inexpensive, easily repeatable, and does not require tissue phantoms.

METHODS

The test protocol is based on in-air reverberation images. The software test tool generates uniformity- and reverberation profiles to monitor system sensitivities and signal uniformities that provide a sensitive analysis of transducer status. Whenever a transducer was suspected to be damaged validating tests were conducted using the Sonora FirstCall test system. A total of 21 transducers from five ultrasound scanner systems were included in the study. Tests were performed every other month for a total of 5 years.

RESULTS

Each transducer was tested on average 11.7 times. Testing a transducer required a total of 2.75 hours annually. The ultrasound quality assurance test protocol detected an average annual fail rate of 10.7%. The test protocol provides a reliable approach for monitoring transducer lens status of clinically used ultrasound transducers.

CONCLUSION

The ultrasound quality assurance test protocol may potentially find deviations in diagnostic quality before being noticed by the clinicians. Thus, the ultrasound quality assurance test protocol holds the ability to reduce the risk of undiscovered image quality degradation hereby minimizing the danger of diagnostic errors.

摘要

目的

实施、执行并评估一种基于软件、与用户无关、成本低廉、易于重复且无需组织体模的临床超声换能器质量保证测试方案。

方法

该测试方案基于空气中的混响图像。软件测试工具生成均匀性和混响剖面图,以监测系统灵敏度和信号均匀性,从而对换能器状态进行灵敏分析。每当怀疑换能器损坏时,就使用索诺拉FirstCall测试系统进行验证测试。本研究共纳入了来自五个超声扫描仪系统的21个换能器。每隔一个月进行一次测试,共进行5年。

结果

每个换能器平均测试11.7次。每年测试一个换能器总共需要2.75小时。超声质量保证测试方案检测到的年平均故障率为10.7%。该测试方案为监测临床使用的超声换能器的透镜状态提供了一种可靠的方法。

结论

超声质量保证测试方案可能在临床医生注意到之前就发现诊断质量的偏差。因此,超声质量保证测试方案有能力降低未被发现的图像质量下降的风险,从而将诊断错误的危险降至最低。

相似文献

1
Five-year evaluation of a low-cost quality assurance protocol for clinical ultrasound transducers.
Ultrasound. 2023 Feb;31(1):71-78. doi: 10.1177/1742271X221091721. Epub 2022 May 23.
2
The analysis of in-air reverberation patterns from medical ultrasound transducers.
Ultrasound. 2014 Feb;22(1):26-36. doi: 10.1177/1742271X13517693. Epub 2014 Feb 6.
3
Quality assurance in diagnostic ultrasound.
Eur J Radiol. 2011 Nov;80(2):519-25. doi: 10.1016/j.ejrad.2010.11.015. Epub 2010 Dec 8.
4
Image Analysis for Ultrasound Quality Assurance.
Ultrason Imaging. 2021 May;43(3):113-123. doi: 10.1177/0161734621992332. Epub 2021 Feb 15.
5
Automated quality control of ultrasound based on in-air reverberation patterns.
Ultrasound. 2017 Nov;25(4):229-238. doi: 10.1177/1742271X17733145. Epub 2017 Sep 29.
6
Ultrasound Probe Acceptance Testing Using the In-Air Reverberation Pattern.
Ultrasound Med Biol. 2019 Jun;45(6):1475-1482. doi: 10.1016/j.ultrasmedbio.2019.02.001. Epub 2019 Mar 30.
7
Routine Ultrasound Quality Assurance in a Multi-Unit Radiology Department: A Retrospective Evaluation of Transducer Failures.
Ultrasound Med Biol. 2017 Sep;43(9):1930-1937. doi: 10.1016/j.ultrasmedbio.2017.05.009. Epub 2017 Jun 17.
8
A simple uniformity test for ultrasound phased arrays.
Phys Med. 2016 Sep;32(9):1162-6. doi: 10.1016/j.ejmp.2016.08.011. Epub 2016 Aug 18.
10
Phantom-based quality assurance measurements in B-mode ultrasound.
Acta Radiol Short Rep. 2013 Dec 11;2(8):2047981613511967. doi: 10.1177/2047981613511967. eCollection 2013.

本文引用的文献

1
Blinded Comparison between an In-Air Reverberation Method and an Electronic Probe Tester in the Detection of Ultrasound Probe Faults.
Ultrasound Med Biol. 2017 Dec;43(12):2954-2958. doi: 10.1016/j.ultrasmedbio.2017.08.936. Epub 2017 Sep 28.
2
The analysis of in-air reverberation patterns from medical ultrasound transducers.
Ultrasound. 2014 Feb;22(1):26-36. doi: 10.1177/1742271X13517693. Epub 2014 Feb 6.
4
Four-year experience with a clinical ultrasound quality control program.
Ultrasound Med Biol. 2011 Aug;37(8):1350-7. doi: 10.1016/j.ultrasmedbio.2011.05.007. Epub 2011 Jun 16.
5
Quality assurance in diagnostic ultrasound.
Eur J Radiol. 2011 Nov;80(2):519-25. doi: 10.1016/j.ejrad.2010.11.015. Epub 2010 Dec 8.
6
Ultrasound transducer function: annual testing is not sufficient.
Eur J Echocardiogr. 2010 Oct;11(9):801-5. doi: 10.1093/ejechocard/jeq072. Epub 2010 Jun 8.
7
High incidence of defective ultrasound transducers in use in routine clinical practice.
Eur J Echocardiogr. 2009 May;10(3):389-94. doi: 10.1093/ejechocard/jen295. Epub 2008 Oct 22.
8
Objective performance testing and quality assurance of medical ultrasound equipment.
Ultrasound Med Biol. 2007 Mar;33(3):460-71. doi: 10.1016/j.ultrasmedbio.2006.09.006.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验