Suppr超能文献

正颌手术标志的区域体素配准线性精度是多少?

What Is the Linear Accuracy of Regional Voxel-Based Registration for Orthognathic Surgery Landmarks?

机构信息

Assistant Professor, Department of Oral and Maxillofacial Surgery, University of Illinois Chicago, Chicago, IL.

Professor, Department of Oral and Maxillofacial Surgery, Kyungpook National University, Daegu, South Korea.

出版信息

J Oral Maxillofac Surg. 2023 May;81(5):546-556. doi: 10.1016/j.joms.2023.01.015. Epub 2023 Feb 21.

Abstract

PURPOSE

While regional voxel-based registration (R-VBR) has been shown to have excellent reproducibility and angular accuracy, there are limited data on the linear accuracy of R-VBR for common orthognathic surgery landmarks, or on whether angular accuracy correlates with linear accuracy. The purpose of this study was to estimate the linear accuracy of R-VBR for several skeletal landmarks commonly used in orthognathic surgical planning, and to measure the correlation between angular and linear discrepancies.

MATERIALS AND METHODS

This is a retrospective cross-sectional study of consecutive patients treated at a single center with nonsegmental LeFort I and bilateral sagittal split osteotomy surgery from January 2019 to November 2020. Cone beam computed tomography at the preoperative (T0) and immediate postoperative (T1) stages were analyzed to measure the postoperative positional changes of 11 orthognathic landmarks in 4 regions of interest (ROI) using R-VBR performed twice by two examiners. Pairwise correlation analysis and canonical correlation analysis were performed for the angular discrepancies (primary predictor variable) and the linear discrepancies (primary outcome variable) to measure the correlation between the two.

RESULTS

In cone beam computed tomography analysis of 28 eligible subjects (16 males, 12 females; mean age 18.9 years, range 15 to 25), the mean absolute (MA) angular discrepancies ranged from 0.15° to 0.55°, while the corresponding MA linear discrepancies ranged from 0.05 to 0.41 mm. There was a strong correlation between angular and linear discrepancies that was statistically significant (P = .001 to .04, Spearman's rank correlation coefficient 0.38 to 0.87).

CONCLUSIONS

For nonsegmental LeFort I osteotomies and bilateral sagittal split osteotomy, R-VBR has excellent linear accuracy within a single voxel size (0.3 mm) for commonly used orthognathic landmarks in the maxillary and distal mandibular ROI. The MA linear discrepancy for the proximal mandibular segment ROI was greater than a single voxel size, with a maximum of 0.41 mm.

摘要

目的

虽然区域体素配准(R-VBR)已被证明具有出色的可重复性和角度精度,但关于 R-VBR 在常见正颌手术标志中的线性精度的数据有限,或者关于角度精度与线性精度是否相关的数据也有限。本研究的目的是评估 R-VBR 在正颌手术规划中常用的几种骨骼标志中的线性精度,并测量角度和线性差异之间的相关性。

材料和方法

这是一项回顾性的横断面研究,纳入了 2019 年 1 月至 2020 年 11 月在一家中心接受非节段性 LeFort I 和双侧矢状劈开截骨术治疗的连续患者。使用由两位检查者两次执行的 R-VBR 分析术前(T0)和即刻术后(T1)阶段的锥形束计算机断层扫描,以测量 4 个感兴趣区域(ROI)中 11 个正颌标志的术后位置变化。进行成对相关分析和典型相关分析,以测量两种角度差异(主要预测变量)和线性差异(主要结果变量)之间的相关性。

结果

在对 28 名符合条件的受试者(16 名男性,12 名女性;平均年龄 18.9 岁,范围 15 至 25 岁)的锥形束计算机断层扫描分析中,平均绝对(MA)角度差异范围为 0.15°至 0.55°,而相应的 MA 线性差异范围为 0.05 至 0.41 毫米。角度和线性差异之间存在很强的相关性,且具有统计学意义(P =.001 至.04,Spearman 等级相关系数为 0.38 至 0.87)。

结论

对于非节段性 LeFort I 截骨术和双侧矢状劈开截骨术,R-VBR 在单个体素大小(0.3 毫米)内具有出色的线性精度,适用于上颌和下颌远侧 ROI 中的常用正颌标志。下颌近侧段 ROI 的 MA 线性差异大于单个体素大小,最大为 0.41 毫米。