Cornea. 2023 Sep 1;42(9):1104-1109. doi: 10.1097/ICO.0000000000003158. Epub 2023 Feb 16.
The aim of this study was to compare the evolution of corneal pachymetry after customized corneal crosslinking (CXL) between Scheimpflug-based and optical coherence-based corneal tomography (OCT).
In this retrospective study, central corneal thickness (CCT), thinnest corneal thickness, and epithelial thickness of 33 eyes of 33 patients with keratoconus were measured preoperatively and 1, 3, and 12 months after customized CXL using the Pentacam HR and the MS-39. The mean pachymetry values of measurements were compared with a paired sample t test. Bland-Altman plots and 95% limits of agreement (LoA) were used to assess the agreement between the measurements of the 2 devices.
The mean age of the participants was 29.7 ± 11.4 years. At baseline, the mean CCT measurements were equal with Pentacam HR (478.30 ± 36.77 μm) and MS-39 (478.46 ± 38.01 μm). After CXL, CCT obtained by Pentacam HR was 460.65 ± 38.69 μm, 464.65 ± 44.45 μm, and 476.77 ± 39.85 μm, and by MS-39 was 478.18 ± 39.50 μm, 472.89 ± 40.92 μm, and 479.51 ± 39.20 μm at 1, 3, and 12 months, respectively. Pentacam HR measured significantly lower CCT ( P < 0.05) at months 1 and 3 after CXL. The agreement was smallest between both devices at month 1 (95% LoA -59 to 24 μm) followed by month 3 (95% LoA: -41 to 23 μm). Epithelial thickness, measured with OCT alone, increased significantly at 1 month and regained preoperative levels at 3 months and thereafter.
After CXL, corneal pachymetry significantly differs between OCT-based and Scheimpflug-based corneal tomography. Pentacam HR seems to underestimate pachymetry when haze is present.
本研究旨在比较基于 Scheimpflug 的角膜光学相干断层扫描(OCT)和基于光学相干的角膜交联(CXL)后角膜生物力学参数的变化。
回顾性研究纳入 33 例(33 只眼)圆锥角膜患者,使用 Pentacam HR 和 MS-39 分别于术前和术后 1、3、12 个月测量中央角膜厚度(CCT)、最薄角膜厚度和角膜上皮厚度。采用配对样本 t 检验比较两种仪器的平均角膜厚度值。Bland-Altman 图和 95%一致性界限(LoA)用于评估两种仪器测量值的一致性。
参与者的平均年龄为 29.7 ± 11.4 岁。在基线时,Pentacam HR(478.30 ± 36.77 μm)和 MS-39(478.46 ± 38.01 μm)的 CCT 测量值相等。CXL 后,Pentacam HR 的 CCT 分别为 460.65 ± 38.69 μm、464.65 ± 44.45 μm 和 476.77 ± 39.85 μm,MS-39 的 CCT 分别为 478.18 ± 39.50 μm、472.89 ± 40.92 μm 和 479.51 ± 39.20 μm,分别在术后 1、3 和 12 个月时测量。CXL 后 1 和 3 个月,Pentacam HR 测量的 CCT 显著较低(P < 0.05)。两种设备之间的一致性在术后 1 个月时最小(95%LoA:-59 至 24 μm),随后在术后 3 个月(95%LoA:-41 至 23 μm)时最小。单独使用 OCT 测量的角膜上皮厚度在术后 1 个月时显著增加,并在术后 3 个月和之后恢复术前水平。
CXL 后,OCT 与基于 Scheimpflug 的角膜断层扫描之间的角膜生物力学参数差异显著。当存在混浊时,Pentacam HR 似乎低估了角膜厚度。