高分辨率Scheimpflug 成像、傅里叶域光学相干断层扫描和超声角膜测厚仪测量中央角膜厚度。

Measurement of central corneal thickness by high-resolution Scheimpflug imaging, Fourier-domain optical coherence tomography and ultrasound pachymetry.

机构信息

School of Optometry and Ophthalmology and Eye Hospital, Wenzhou Medical College, Wenzhou, Zhejiang, China.

出版信息

Acta Ophthalmol. 2012 Aug;90(5):449-55. doi: 10.1111/j.1755-3768.2010.01947.x. Epub 2010 Jun 18.

Abstract

PURPOSE

To compare the repeatability and reproducibility of central corneal thickness (CCT) measurements by high-resolution (HR) rotating Scheimpflug imaging and Fourier-domain optical coherence tomography (FD-OCT). CCT measurements were compared to those determined by ultrasound pachymetry (UP).

METHODS

In 35 healthy eyes, intra-observer repeatability for HR Scheimpflug (Pentacam) and FD-OCT (RTVue) systems was determined in consecutive images taken by an observer in the shortest time possible. Imaging was repeated again by a second observer to evaluate inter-observer reproducibility. The CCT measurements were compared among Scheimpflug, FD-OCT and UP images.

RESULTS

Mean coefficients of repeatability were 0.48% for Scheimpflug and 0.26% for FD-OCT. For Scheimpflug, the coefficient of inter-operator reproducibility was 0.87%. For FD-OCT, the coefficient of inter-operator reproducibility was 0.45%. The CCT measurements by Scheimpflug, OCT and UP images were (mean ± standard deviation) 521.7 ± 27.6 μm, 510.8 ± 28.6 μm and 516.5 ± 27.6 μm, respectively. The differences between instruments were statistically significant. The 95% limits of agreement in CCT were -0.7 to 22.5 μm for Pentacam-OCT, -13.4 to 24.0 μm for Pentacam-UP and -26.7 to 15.4 μm for OCT-UP. There was a high degree of correlation between CCT measured by all 3 methods.

CONCLUSION

Noncontact measurements of CCT with HR Scheimpflug and FD-OCT systems yielded excellent repeatability and reproducibility and can be used interchangeably. Although both devices were comparable with UP; in clinical practice, the measurements acquired by optical modalities are not directly interchangeable with UP measurements.

摘要

目的

比较高分辨率(HR)旋转式 Scheimpflug 成像和傅里叶域光学相干断层扫描(FD-OCT)测量中央角膜厚度(CCT)的可重复性和再现性。将 CCT 测量值与超声角膜测厚术(UP)确定的值进行比较。

方法

在 35 只健康眼中,观察者以尽可能短的时间连续拍摄 HR Scheimpflug(Pentacam)和 FD-OCT(RTVue)系统的图像,以确定观察者内的重复性。由第二位观察者再次进行成像,以评估观察者间的再现性。Scheimpflug、FD-OCT 和 UP 图像之间比较 CCT 测量值。

结果

Scheimpflug 的重复性平均系数为 0.48%,FD-OCT 的重复性平均系数为 0.26%。Scheimpflug 的操作者间可重复性系数为 0.87%。FD-OCT 的操作者间可重复性系数为 0.45%。Scheimpflug、OCT 和 UP 图像的 CCT 测量值(平均值±标准差)分别为 521.7±27.6μm、510.8±28.6μm 和 516.5±27.6μm。仪器之间的差异具有统计学意义。Pentacam-OCT 的 CCT 95%一致性界限为-0.7 至 22.5μm,Pentacam-UP 的 CCT 95%一致性界限为-13.4 至 24.0μm,OCT-UP 的 CCT 95%一致性界限为-26.7 至 15.4μm。所有 3 种方法测量的 CCT 之间具有高度相关性。

结论

HR Scheimpflug 和 FD-OCT 系统的 CCT 非接触测量具有极好的重复性和再现性,可以互换使用。尽管这两种设备与 UP 相比都具有可比性;但在临床实践中,光学模态获取的测量值与 UP 测量值不能直接互换。