J Med Life. 2023 Jan;16(1):10-15. doi: 10.25122/jml-2022-0269.
The purpose of single-use flexible ureteroscopes (su-fURS) was to overcome the limitations of conventional reusable ureteroscopes in terms of maneuverability and maintenance. We aimed to perform a systematic literature review on available su-fURS performance conventional reusable fURS focusing on clinical data. A systematic research using Pubmed was performed evaluating single-use fURS and reusable fURS in urinary tract stone disease, including prospective assessments and case series. This review aimed to provide an overview of single-use and disposable flexible ureteroscopes and to examine and compare their capabilities (deflection, irrigation, optical properties). We included 11 studies, where the single-use fURS were compared to the reusable fURS. The studies with single-use ureteroscopes included data on LithoVue (Boston Scientific), The Uscope UE3022 (Pusen, Zhuhai, China), NeoFlex-Flexible, (Neoscope Inc San Jose, CA), 23 YC-FR-A (Shaogang). For reusable ureteroscopes, data were included on three models, two digital (Karl Storz Flex-XC and Olympus URF-Vo) and one fiber optic (Wolf-Cobra). There were no significant differences in stone-free rate, procedure duration, or functional capabilities between single-use fURS and reusable fURS. The systematic literature review analyzed operative time, functional capabilities, stone-free rates, and postoperative complications of the ureteroscopes, and a special chapter about renal abnormalities to emphasize that they are a good choice having a high proportion of stone-free rates and few risks, particularly in treating difficult-to-access calculi. Single-use fURS demonstrate a comparable efficacy with reusable fURS in resolving renal lithiasis. Further studies on clinical efficacy are needed to determine whether single-use fURS will reliably replace its reusable counterpart.
一次性使用软性输尿管镜(su-fURS)的目的是克服传统可重复使用输尿管镜在操作性和维护方面的局限性。我们旨在对现有的 su-fURS 性能进行系统文献综述,重点关注临床数据,与传统可重复使用的 fURS 进行比较。使用 Pubmed 进行了系统研究,评估了一次性使用 fURS 和可重复使用 fURS 在尿路结石病中的应用,包括前瞻性评估和病例系列研究。本综述旨在提供一次性使用和一次性使用软性输尿管镜的概述,并检查和比较它们的功能(弯曲度、冲洗、光学性能)。我们纳入了 11 项研究,其中将一次性输尿管镜与可重复使用输尿管镜进行了比较。使用一次性输尿管镜的研究包括 LithoVue(Boston Scientific)、The Uscope UE3022(Pusen,珠海,中国)、NeoFlex-Flexible(Neoscope Inc San Jose,CA)、23 YC-FR-A(Shaogang)的数据。对于可重复使用的输尿管镜,纳入了三种型号的数据,两种数字(Karl Storz Flex-XC 和 Olympus URF-Vo)和一种光纤(Wolf-Cobra)。一次性使用 fURS 和可重复使用 fURS 之间在无结石率、手术时间或功能能力方面没有显著差异。系统文献综述分析了输尿管镜的手术时间、功能能力、无结石率和术后并发症,以及一个关于肾脏异常的特殊章节,以强调它们是一个很好的选择,具有较高的无结石率和较低的风险,特别是在治疗难以到达的结石。一次性使用 fURS 在解决肾结石方面与可重复使用 fURS 具有相当的疗效。需要进一步的临床疗效研究来确定一次性使用 fURS 是否能可靠地替代可重复使用的输尿管镜。