Kamei Daigo, Nagano Masashi, Takagaki Takashi, Sakamoto Tomotaka, Tsuchiya Ken
Department of Blood Purification, Kidney Center, Tokyo Women's Medical University, 8-1 Kawadacho, Shinjyuku-ku, Tokyo, 162-8666, Japan.
Nerima Sakuradai Clinic, 4-11-9 Toyotamakita, Nerima-ku, Tokyo, 176-0012, Japan.
Heliyon. 2023 Feb 20;9(3):e13896. doi: 10.1016/j.heliyon.2023.e13896. eCollection 2023 Mar.
Hepcidin-25 is an iron regulatory factor that plays an important role in anemia in patients with chronic kidney disease. Although liquid chromatography/tandem mass spectroscopy (LC-MS/MS) is the gold standard for measuring hepcidin-25 concentrations, results cannot be obtained immediately at clinical sites. In contrast, the latex immunoassay (LIA) can be performed using general clinical laboratory equipment, and the results can be obtained quickly. The aim of this study was to evaluate hepcidin-25 concentrations obtained by LC-MS/MS and a novel LIA method and compare the two methods.
Hepcidin-25 was measured by LIA and LC-MS/MS in 182 hemodialysis patients. A hepcidin-25-specific reagent and an automatic analyzer were used for LIA, and a commercially available system was used for LC-MS/MS. The Passing-Bablok regression analysis method was used.
On Passing-Bablok regression analysis, the slope was 1.000, with an intercept of 0.359. Very strong associations were obtained, and the measured values were almost identical.
The hepcidin-25 concentrations measured using LIA and those measured by LC-MS/MS were significantly correlated. LIA can be performed using general clinical examination equipment and has a higher throughput than LC-MS/MS. Therefore, measurement of hepcidin-25 concentrations by LIA can be useful for routine laboratory testing.
铁调素-25是一种铁调节因子,在慢性肾脏病患者的贫血中起重要作用。尽管液相色谱/串联质谱法(LC-MS/MS)是测定铁调素-25浓度的金标准,但临床现场无法立即获得结果。相比之下,乳胶免疫测定法(LIA)可使用普通临床实验室设备进行,且能快速获得结果。本研究的目的是评估通过LC-MS/MS和一种新型LIA方法获得的铁调素-25浓度,并比较这两种方法。
对182例血液透析患者采用LIA和LC-MS/MS测定铁调素-25。LIA使用铁调素-25特异性试剂和自动分析仪,LC-MS/MS使用市售系统。采用Passing-Bablok回归分析法。
在Passing-Bablok回归分析中,斜率为1.000,截距为0.359。获得了很强的相关性,测量值几乎相同。
使用LIA测定的铁调素-25浓度与通过LC-MS/MS测定的浓度显著相关。LIA可使用普通临床检查设备进行,且通量高于LC-MS/MS。因此,通过LIA测定铁调素-25浓度可用于常规实验室检测。