Tensor velocity imaging (TVI) performance with a row-column probe was assessed for constant flow in a straight vessel phantom and pulsatile flow in a carotid artery phantom. TVI, i.e., estimating the 3-D velocity vector as a function of time and spatial position, was performed using the transverse oscillation cross-correlation estimator, and the flow was acquired with a Vermon 128+128 row-column array probe connected to a Verasonics 256 research scanner. The emission sequence used 16 emissions per image, and a TVI volume rate of 234 Hz was obtained for a pulse repetition frequency (f) of 15 kHz. The TVI was validated by comparing estimates of the flow rate through several cross-sections with the flow rate set by the pump. For the constant 8 mL/s flow in the straight vessel phantom with relative estimator bias (RB) and standards deviation (RSD) was found in the range of -2.18% to 0.55% and 4.58% to 2.48% in measurements performed with an f of 15, 10, 8, and 5 kHz. The pulsatile flow in the carotid artery phantom the was set to an average flow rate of 2.44 mL/s, and the flow was acquired with an f of 15, 10, and 8 kHz. The pulsatile flow was estimated from two measurement sites: one at a straight section of the artery and one at the bifurcation. In the straight section, the estimator predicted the average flow rate with an RB value ranging from -7.99% to 0.10% and an RSD value ranging from 10.76% to 6.97%. At the bifurcation, RB and RSD values were between -7.47% to 2.02% and 14.46% to 8.89%. This demonstrates that an RCA with 128 receive elements can accurately capture the flow rate through any cross-section at a high sampling rate.
采用行列探头评估了恒流直血管体模和脉动流颈动脉体模中张量速度成像(TVI)的性能。TVI 即通过横向振动互相关估计器来估计作为时间和空间位置函数的 3-D 速度向量,并使用连接到 Verasonics 256 研究扫描仪的 Vermon 128+128 行列探头获取流量。发射序列每幅图像使用 16 次发射,脉冲重复频率(f)为 15 kHz 时,可获得 234 Hz 的 TVI 体率。通过将几个横截面的流量估计值与泵设定的流量进行比较,对 TVI 进行了验证。在直血管体模中,当 f 为 15、10、8 和 5 kHz 时,对于恒定的 8 mL/s 流量,相对估计偏差(RB)和标准偏差(RSD)的范围分别为-2.18%至 0.55%和 4.58%至 2.48%。在颈动脉体模中,脉动流的平均流量设定为 2.44 mL/s,f 为 15、10 和 8 kHz 时采集流量。从两个测量位置估计脉动流:一个在动脉的直段,一个在分叉处。在直段,估计器以 RB 值范围从-7.99%至 0.10%和 RSD 值范围从 10.76%至 6.97%预测平均流量。在分叉处,RB 和 RSD 值在-7.47%至 2.02%和 14.46%至 8.89%之间。这表明,具有 128 个接收元件的 RCA 可以以高采样率准确捕获任何横截面的流量。