Department of Orthopaedic Surgery, Shimura Hospital.
Okimoto Clinic.
J Nippon Med Sch. 2023;90(1):103-110. doi: 10.1272/jnms.JNMS.2023_90-116.
In vivo three-dimensional (3D) motion under weight-bearing conditions was analyzed postoperatively in medial pivot cruciate-substituting (CS) knee systems with fixed and mobile inserts.
Tibiofemoral knee kinematics during squatting were captured with X-ray fluoroscopy for 4 patients in each cohort. The 3D motion of implants was analyzed with KneeMotion motion analysis software (LEXI Corporation; Tokyo, Japan). In addition, anterior-posterior (AP) movement of the distal-most points and the angle of axial rotation of the femoral component on the tibial component were assessed in both cohorts.
Mean AP movement of the femoral component on the tibial component was 3.8±0.5 mm on the medial side and 9.5±0.5 mm on the lateral side in the cohort with fixed prostheses and 5.9±2.1 mm on the medial side and 10.0±2.5 mm on the lateral side in the cohort with mobile prostheses. The mean angle of axial rotation of the femoral component on the tibial component was 14.4±1.1 degrees and 8.2±2.7 degrees of external rotation for fixed knees and mobile knees, respectively.
Postoperative motion analysis confirmed that fixed and mobile CS implants, which have a similar design, guided medial pivot motion under weight-bearing conditions. However, motion differed between these implant types after mid-flexion: bicondylar rollback after medial pivot motion was noted in the mobile cohort.
本研究分析了在负重条件下,内侧枢轴置换式(CS)膝关节系统中固定和活动衬垫术后的体内三维(3D)运动。
4 例患者分别纳入两组,在蹲位时采用 X 射线透视采集胫股关节运动学数据。使用 KneeMotion 运动分析软件(LEXI 公司;东京,日本)分析植入物的 3D 运动。此外,还评估了两组中股骨组件在胫骨组件上的近侧最远端的前后(AP)移动和轴向旋转角度。
固定假体组中股骨组件在胫骨组件上的平均 AP 移动为内侧 3.8±0.5mm,外侧 9.5±0.5mm;活动假体组中股骨组件在胫骨组件上的平均 AP 移动为内侧 5.9±2.1mm,外侧 10.0±2.5mm。股骨组件在胫骨组件上的平均轴向旋转角度为固定膝 14.4±1.1 度,活动膝 8.2±2.7 度的外旋。
术后运动分析证实,具有相似设计的固定和活动 CS 植入物在负重条件下引导内侧枢轴运动。然而,在过屈位时,这两种植入物类型的运动存在差异:在活动组中观察到内侧枢轴运动后的双髁后退。