Shaw C G, Plewes D B
Department of Radiology, University of Rochester Medical Center, New York 14642.
Med Phys. 1987 Nov-Dec;14(6):956-67. doi: 10.1118/1.595975.
Dual-energy subtraction imaging allows tissue and bone structures to be separated from each other and attenuating thicknesses measured. Potential applications include chest imaging, bone mineral measurement, angiography, and mammography. However, intrinsic to most x-ray detectors is the acceptance of scattered radiation as part of the image signal. Added to that is the veiling glare component when an image intensifier is used. Together, they result in erroneous transmission measurement and degrade the accuracy of energy subtraction processing. In this paper, the effects of scattered radiation and veiling glare on energy subtraction images are examined theoretically. A model is derived and used to compute the effects on the thickness signals, image contrast, and image noise as a function of the scatter glare to primary ratios. The ratios were measured on a point-by-point basis for a Rando chest phantom. For 96% of the image field studied, the thickness signals may be subject to an error ranging from 0 to -22.5 cm for tissue and 0 to 5.2 cm for bone. The image contrast in the tissue image may be reduced by a factor ranging from 1 to 59. The percentage of the uncanceled bone signals ranges from -52% to 52%. The contrast-to-noise ratio may be reduced by a factor ranging from 1 to 18.
双能减影成像可使组织结构与骨骼结构相互分离,并测量衰减厚度。潜在应用包括胸部成像、骨矿物质测量、血管造影和乳腺摄影。然而,大多数X射线探测器的固有特性是将散射辐射作为图像信号的一部分接收。此外,使用图像增强器时还存在蒙罩眩光成分。它们共同导致传输测量错误,并降低能量减影处理的准确性。本文从理论上研究了散射辐射和蒙罩眩光对能量减影图像的影响。推导了一个模型,并用于计算作为散射眩光与原发射线比率函数的对厚度信号、图像对比度和图像噪声的影响。对Rando胸部模型逐点测量了这些比率。在所研究的96%的图像区域中,组织的厚度信号误差范围可能为0至-22.5厘米,骨骼的厚度信号误差范围可能为0至5.2厘米。组织图像中的图像对比度可能降低1至59倍。未消除的骨骼信号百分比范围为-52%至52%。对比度噪声比可能降低1至18倍。