[Evaluation of Plasma Concentrations of Mycophenolic Acid in Renal Transplant Patients Using LM1010 High-performance Liquid Chromatography].

作者信息

Akamine Yumiko, Matsushita Miyuki, Morikawa Satoru, Miura Masatomo

机构信息

Department of Pharmacy, Akita University Hospital.

Department of Service Promotion, Hitachi High-Tech Science Corporation.

出版信息

Yakugaku Zasshi. 2023;143(4):377-383. doi: 10.1248/yakushi.22-00182.

Abstract

Plasma concentrations of mycophenolic acid (MPA), an immunosuppressive agent, have been measured in clinical settings using immunoassay methods or HPLC. However, immunoassay methods show cross-reactivity with metabolites of MPA glucuronide. Recently, the LM1010 high-performance liquid chromatography instrument was approved as a new general medical device. In this study, we compared the results of MPA plasma concentrations analyzed using the LM1010 method and the previously described HPLC method. Plasma samples obtained from 100 renal transplant patients (32 women and 68 men) were evaluated using both HPLC instruments. Deming regression analyses showed a very high correlation between the two instruments, with a slope of 0.9892 and an intercept of 0.0235 µg/mL (r=0.982). Bland-Altman analysis showed an average of -0.0012 µg/mL between the LM1010 method and the previously described HPLC method. For the LM1010 method, the total run time for MPA analysis was 7 min, and the analytical time was short; however, the extraction recovery when using a spin column was extremely low for frozen plasma samples stored at -20°C for 1 month, and the volume required for the assay (150 µL) could not be collected. Thus, for the LM1010 method, analysis using fresh plasma samples was optimal. Overall, our findings showed that the LM1010 method was a rapid, accurate HPLC assay for MPA analysis and could be used in clinical practice for routine monitoring of MPA in fresh plasma samples.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验