Smith R M, Hurdley T G, Gill R, Moffat A C
J Chromatogr. 1986 Mar 14;355(1):75-85. doi: 10.1016/s0021-9673(01)97304-2.
The separation of the local anaesthetic drugs by reversed-phase high-performance liquid chromatography using a Hypersil ODS column has been examined in order to identify critical factors in the establishment of a retention database for interlaboratory comparisons. An eluent of methanol-aqueous orthophosphoric acid (15:85, v/v) (pH 2.5) containing 0.7% hexylamine has been used. The effects of small changes in the eluent composition, pH, temperature and the use of different column packing materials on the retentions of the analytes have been studied. The use of capacity factors, relative capacity factors and retention indices based on the alkylarylketone scale have been compared.
为了确定建立用于实验室间比较的保留数据库的关键因素,研究了使用Hypersil ODS柱通过反相高效液相色谱法分离局部麻醉药的情况。使用了含有0.7%己胺的甲醇 - 正磷酸水溶液(15:85,v/v)(pH 2.5)作为洗脱液。研究了洗脱液组成、pH值、温度的微小变化以及使用不同柱填料对分析物保留的影响。比较了容量因子、相对容量因子以及基于烷基芳基酮标度的保留指数的使用情况。