Puzzling interference in the Siemens tacrolimus assay in a renal transplant patient: A case report.

机构信息

Department of Laboratory Medicine, University Clinical Hospital of Santiago de Compostela, 15706 Santiago de Compostela, Spain.

Department of Laboratory Medicine, University Clinical Hospital of Santiago de Compostela, 15706 Santiago de Compostela, Spain.

出版信息

Clin Chim Acta. 2023 Mar 15;543:117327. doi: 10.1016/j.cca.2023.117327. Epub 2023 Apr 2.

Abstract

BACKGROUND

Therapeutic drug monitoring is a key tool for optimizing tacrolimus therapy in transplant recipients. The modified ACMIA assay from Siemens Healthcare Diagnostics is an immunoassay commonly used for tacrolimus monitoring, but it is not known whether this assay is resistant to interference from endogenous substances in real-world use.

OBJECTIVE

To describe a case of unexpected interference in tacrolimus monitoring using the modified ACMIA method in a kidney transplant recipient, and to highlight the importance of careful interpretation of laboratory results and effective communication with clinicians in optimizing patient care.

CASE DESCRIPTION

This case report describes a significant interference in the monitoring of tacrolimus in a kidney transplant recipient using the new ACMIA method. In this case, when aberrant results for tacrolimus were found using the new ACMIA method, they were re-analyzed using the CMIA method from Abbott. The presence of positive ANCA-MPO autoantibodies was found to be the most likely cause of the interference after an extensive workup.

CONCLUSIONS

This is the first report of major interference with the modified ACMIA tacrolimus method and emphasizes the importance of proper interpretation of laboratory results and effective communication with clinicians in optimizing patient care.

摘要

背景

治疗药物监测是优化移植受者他克莫司治疗的重要工具。西门子医疗诊断公司的改良 ACMIA 检测法是一种常用于他克莫司监测的免疫检测法,但尚不清楚该检测法在实际应用中是否能抵抗内源性物质的干扰。

目的

描述一例使用改良 ACMIA 法在肾移植受者中进行他克莫司监测时出现意外干扰的病例,强调在优化患者治疗时仔细解释实验室结果和与临床医生有效沟通的重要性。

病例描述

本病例报告描述了一例新 ACMIA 法在肾移植受者中监测他克莫司时出现显著干扰的病例。在这种情况下,当使用新 ACMIA 法发现他克莫司的结果异常时,使用雅培的 CMIA 法重新进行了分析。经过广泛的检查,发现存在阳性 ANCA-MPO 自身抗体,这是最可能导致干扰的原因。

结论

这是首例关于改良 ACMIA 他克莫司方法出现重大干扰的报告,强调了正确解释实验室结果和与临床医生有效沟通在优化患者治疗中的重要性。