Aust Endod J. 2023 Sep;49 Suppl 1:419-425. doi: 10.1111/aej.12758. Epub 2023 Apr 10.
The effect of tooth structure subjected to ionising radiation on oxygen saturation levels (SpO ) measured by pulse oximetry was investigated. Fifty-four extracted human third molars were sectioned mesiodistally. Pulse oximeter diodes were placed in prefabricated silicone moulds, secured on a vise with tooth surfaces placed between the diodes and the participant's finger, and SpO levels were measured. After that, the tooth surfaces were subjected to ionising radiation in fractionated doses of up to 70 Gy, and SpO was measured according to the protocol described. The Kolmogorov-Smirnov test was used for the analysis of normality of the quantitative variables, and Student's t-test compared mean SpO levels before and after irradiation. Mean SpO after irradiation was 94.6 ± 0.9%, without a statistically significant difference from SpO before irradiation, which was 94.7 ± 0.6%. Irradiation of the tooth structure did not interfere with SpO levels measured by pulse oximetry.
本研究旨在探讨经电离辐射后的牙体结构对脉搏血氧饱和度(SpO )的影响。将 54 颗人第三磨牙沿近远中向纵剖,将脉搏血氧饱和度测定仪的二极管置于预制硅树脂模具中,用虎钳固定,将牙面置于二极管和被检者手指之间,测量 SpO 水平。之后,对牙面进行分割剂量达 70Gy 的电离辐射,按照描述的方案测量 SpO 。采用 Kolmogorov-Smirnov 检验对定量变量进行正态性分析,采用 Student's t 检验比较照射前后 SpO 的均值。照射后 SpO 均值为 94.6±0.9%,与照射前的 SpO 均值 94.7±0.6%相比,无统计学差异。牙体结构的照射不干扰脉搏血氧饱和度测定仪所测的 SpO 水平。