Abstract

BACKGROUND

Customized computer-aided-design/computer-aided-manufacturing (CAD/CAM) titanium meshes have been adopted for alveolar bone augmentation. But the inaccuracies between planned and created bone volume/contour are quite common, and the surgical placement of the customized mesh was considered as the first critical factor. However, the evaluation of surgical placement accuracy of customized mesh is currently lacking.

PURPOSE

The aim of this study was to evaluate the accuracy of the surgical placement of customized meshes.

METHODS

A total of 30 cases, 20 without the screws-position-guided template and 10 with the screws-position-guided template, were included in this study. The cone beam CT (CBCT) data sets of pre- and postoperative were converted into 3D models and digitally aligned. Then the actual placement of customized mesh and retainer titanium screws was compared to the virtual one to assess the surgical placement accuracy of customized mesh. At least 6 months after surgery, a new CBCT was taken and converted into 3D models. Planned bone volume, created bone volume, vertical bone augmentation, healing complications rate, pseudo-periosteum rate, exposure rate, and infection rate were all evaluated.

RESULTS

The 3D digital reconstruction/registration analysis showed that the average difference between actual placement and planned one of customized mesh in positive and negative directions was 2.69 ± 0.70 mm and -1.41 ± 0.90 mm, respectively, without the screws-position-guided template. And the mean difference values between the actual and planned placement of the screws on the X and Y axes were 0.74 ± 0.85 mm and 0.89 ± 0.84 mm. In contrast, with the screws-position-guided template, the results were 2.38 ± 0.69 mm and -1.30 ± 1.13 mm. Accordingly, the mean difference values of screws were 0.76 ± 0.84 mm and 0.94 ± 0.72 mm. There was no statistical difference between the two groups, and the noninferiority of the control group compared to the test group was also confirmed by the comparative analysis.

CONCLUSION

It can be concluded that there is a certain deviation between the planned surgical placement and actual one of customized mesh, and using screws-position-guided template is of limited help for its accurate placement. Further research is needed to achieve precise surgical placement of the customized mesh to achieve precise alveolar bone augmentation.

摘要

背景

定制的计算机辅助设计/计算机辅助制造(CAD/CAM)钛网已被用于牙槽骨增强。但是,计划和创建的骨量/轮廓之间的不准确性很常见,并且定制网的手术放置被认为是第一个关键因素。然而,目前还缺乏对定制网手术放置准确性的评估。

目的

本研究旨在评估定制网手术放置的准确性。

方法

本研究共纳入 30 例患者,其中 20 例无螺钉定位引导模板,10 例有螺钉定位引导模板。术前和术后的锥形束 CT(CBCT)数据集被转换为 3D 模型并进行数字对齐。然后将定制网的实际放置位置与保留钛钉的虚拟位置进行比较,以评估定制网的手术放置准确性。术后至少 6 个月,进行新的 CBCT 检查并转换为 3D 模型。评估计划骨量、创建骨量、垂直骨增量、愈合并发症发生率、假性骨膜率、暴露率和感染率。

结果

3D 数字重建/配准分析显示,在无螺钉定位引导模板的情况下,定制网的实际放置位置与计划位置在正、负方向上的平均差异分别为 2.69±0.70mm 和-1.41±0.90mm。而在 X 和 Y 轴上,螺钉实际和计划放置位置的平均差值分别为 0.74±0.85mm 和 0.89±0.84mm。相比之下,使用螺钉定位引导模板时,结果分别为 2.38±0.69mm 和-1.30±1.13mm。相应地,螺钉的平均差值分别为 0.76±0.84mm 和 0.94±0.72mm。两组之间无统计学差异,通过对比分析也证实了对照组与试验组的非劣效性。

结论

可以得出结论,定制网的计划手术放置与实际放置之间存在一定偏差,使用螺钉定位引导模板对其精确定位帮助有限。需要进一步研究以实现定制网的精确手术放置,从而实现精确的牙槽骨增强。