Nucl Med Commun. 2023 Jul 1;44(7):596-603. doi: 10.1097/MNM.0000000000001697. Epub 2023 Apr 17.
The purpose of this study was to evaluate how statistical fluctuation in single-photon emission computed tomography (SPECT) images propagate to absorbed dose maps.
SPECT/computed tomography (CT) images of iodine-131 filled phantoms, using different acquisition and processing protocols, were evaluated using STRATOS software to assess the absorbed dose distribution at the voxel level. Absorbed dose values and coefficient of variation (COV) were analyzed for dosimetry based on single time-point SPECT images and time-integrated activities of SPECT sequences with low and high counts.
Considering dosimetry based on a single time-point, the mean absorbed dose was not significantly affected by total counts or reconstruction parameters, but the uniformity of the absorbed dose maps had an almost linear correlation with SPECT noise. When high- and low-count SPECT sequences were used to generate an absorbed dose map, the absorbed dose COV for each of the temporal sequences was slightly lower than the absorbed dose COV based on the single SPECT image with the highest count included in the sequence.
The impact of changes in SPECT counts and reconstruction parameters is almost linear when dosimetry is based on isolated SPECT images, but less pronounced when dosimetry is based on sequential SPECTs.
本研究旨在评估单光子发射计算机断层扫描(SPECT)图像中的统计波动如何传播到吸收剂量图。
使用不同的采集和处理方案,使用 STRATOS 软件对碘-131 填充的体模的 SPECT/CT 图像进行评估,以评估体素水平的吸收剂量分布。基于单点 SPECT 图像和低计数与高计数的 SPECT 序列的时间积分活动,分析吸收剂量值和变异系数(COV)进行剂量测定。
考虑基于单点的剂量测定,总计数或重建参数对平均吸收剂量没有显著影响,但吸收剂量图的均匀性与 SPECT 噪声几乎呈线性相关。当使用高低计数的 SPECT 序列生成吸收剂量图时,每个时间序列的吸收剂量 COV 略低于包含在序列中的最高计数的单个 SPECT 图像的吸收剂量 COV。
当基于孤立的 SPECT 图像进行剂量测定时,SPECT 计数和重建参数的变化几乎呈线性,但当基于连续的 SPECT 进行剂量测定时,其影响则不太明显。