Bath P E, Romberger A, Brown P, Quon D
J Cataract Refract Surg. 1986 May;12(3):262-6. doi: 10.1016/s0886-3350(86)80005-0.
Damage to the intraocular lens (IOL) in Nd:YAG laser posterior capsulotomy is a complication to be avoided. The damage threshold for many of the commercially available IOLs was determined. The power density at the focal point was calculated for FDA approved Nd:YAG lasers. Based on the therapeutic gap concept it is recommended that only power levels below the damage threshold be used in photodiscission posterior capsulotomy to minimize the risk of Nd:YAG laser IOL damage.