Diabetes Technol Ther. 2023 Aug;25(8):538-542. doi: 10.1089/dia.2023.0070. Epub 2023 May 4.
There are limited data on head-to-head performance of Freestyle Libre Pro (FSL-Pro) and blinded Medtronic iPro2 continuous glucose monitoring system in pregnancy. In this prospective observational study, women with hyperglycemia in pregnancy ( = 42) underwent simultaneous FSL-Pro and Medtronic iPro2 sensor insertion and self-monitoring of blood glucose using Contour Plus meter (reference). The overall mean absolute relative difference (MARD) for iPro2 and FSL-Pro systems were 8.0% ± 9.2% and 19.0% ± 12.7%, respectively. At hypoglycemic range, both sensors performed less accurately (MARD: 18.0% and 16.8%, respectively), whereas iPro2 showed higher accuracy at euglycemic (8.2% and 19.3%, respectively) and hyperglycemic (6.8% and 18.0%, respectively) ranges. On Bland-Altman analysis, iPro2 and FSL-Pro underestimated glucose by 0.01 and 1.09 mmol/L, respectively. The ISO criteria were fulfilled for 88.5% and 44.9% of all values, respectively. To conclude, iPro2 was more accurate; however, both sensors demonstrated inaccuracy at hypoglycemic range, highlighting the need for refinements in the current generation of sensors to address this problem.
关于 Freestyle Libre Pro(FSL-Pro)和盲法 Medtronic iPro2 连续血糖监测系统在妊娠中的头对头性能,数据有限。在这项前瞻性观察研究中,患有妊娠性高血糖的女性(n=42)同时插入 FSL-Pro 和 Medtronic iPro2 传感器,并使用 Contour Plus 血糖仪(参考)进行自我血糖监测。iPro2 和 FSL-Pro 系统的总体平均绝对相对差异(MARD)分别为 8.0%±9.2%和 19.0%±12.7%。在低血糖范围内,两种传感器的准确性都较低(MARD:分别为 18.0%和 16.8%),而 iPro2 在正常血糖(MARD:分别为 8.2%和 19.3%)和高血糖(MARD:分别为 6.8%和 18.0%)范围内的准确性更高。在 Bland-Altman 分析中,iPro2 和 FSL-Pro 分别低估了 0.01 和 1.09mmol/L 的血糖值。ISO 标准分别满足 88.5%和 44.9%的所有值。总之,iPro2 更准确;然而,两种传感器在低血糖范围内均表现出不准确性,这凸显了当前一代传感器需要改进以解决这一问题。