La Fontaine R, Stein M A, Graham L S, Winter J
Radiology. 1986 Jul;160(1):255-60. doi: 10.1148/radiology.160.1.3715039.
Cold-lesion contrast resolution was measured as a function of photopeak window position, photon energy, and scatter configuration using a clinical scintillation camera capable of asymmetric photopeak imaging. Two radionuclide solutions, technetium-99m and thallium-201, were studied at five photopeak window positions: the symmetric setting and window settings that were 5%, 10%, 20%, and 30% asymmetrically high (asymmetry was defined here by the intrinsic percentage of count loss relative to the symmetric window setting). Low contrast was examined with small cold-sphere phantoms and large cold-cube phantoms to produce high- and low-scatter conditions, respectively. High contrast was examined with a Hine cylindrical step phantom. Lesion contrast improved markedly with small degrees of window asymmetry but showed less dramatic improvement when positions at the higher degrees of asymmetry were studied. For the high-scatter and low-contrast situations, the maximum improvement in lesion contrast tended to be inversely proportional to lesion size. The maximum improvement in contrast resolution ranged from 10% to 45%, depending on lesion size, scatter configuration, and the radionuclide solution used.
使用能够进行不对称光峰成像的临床闪烁相机,将冷病变对比度分辨率作为光峰窗口位置、光子能量和散射配置的函数进行测量。研究了两种放射性核素溶液,锝-99m和铊-201,在五个光峰窗口位置:对称设置以及分别比对称设置高5%、10%、20%和30%的不对称窗口设置(此处不对称性由相对于对称窗口设置的计数损失固有百分比定义)。使用小冷球模体和大冷立方体模体分别产生高散射和低散射条件来检查低对比度。使用海因圆柱形阶梯模体检查高对比度。病变对比度在窗口不对称程度较小时显著提高,但在研究更高不对称程度的位置时,改善程度较小。对于高散射和低对比度情况,病变对比度的最大改善往往与病变大小成反比。对比度分辨率的最大改善范围为10%至45%,具体取决于病变大小、散射配置和所使用的放射性核素溶液。