圆锥角膜患者角膜交联术后的角膜生物力学变化

Corneal Biomechanical Changes after Corneal Cross-Linking in Patients with Keratoconus.

作者信息

Hamid Adel, Jahadi-Hosseini Hamidreza, Khalili Mohammad Reza, Jahanbani-Ardakani Hamidreza

机构信息

Poostchi Ophthalmology Research Center, Department of Ophthalmology, School of Medicine, Shiraz University of Medical Sciences, Shiraz, Iran.

出版信息

J Curr Ophthalmol. 2023 Apr 29;34(4):409-413. doi: 10.4103/joco.joco_177_21. eCollection 2022 Oct-Dec.

Abstract

PURPOSE

To evaluate the changes in biomechanical properties of the cornea using the Corvis ST device after the treatment of keratoconus patients with ultraviolet-A/riboflavin corneal cross-linking (CXL).

METHODS

Thirty-seven eyes from 37 consecutive patients with progressive keratoconus were included in this prospective observational case series. Corneal biomechanical parameters including the length of the applanated cornea (L1 and L2), corneal movement velocity during applanation (V1 and V2) at the moment of the first and second applanation, deformation amplitude (DA), distance between bending points of the cornea points of the cornea (PD), and concave radius of curvature (R) at the point of the highest concavity were recorded using the Corvis ST at baseline, 3 months, and 1 year after CXL.

RESULTS

The mean age of the patients was 23.27 years (range, 19-31 years). Among CorVis ST corneal biomechanical parameters, L1, DA, PD, and R at the point of the highest concavity did not change significantly. The length of the applanated cornea at the moment of second applanation (L2) showed a significant change 3 months after CXL, but no significant difference was found between the 3-month and 1-year values of this parameter. Corneal movement velocity during applanation (V1 and V2) did not change 3 months after doing CXL, but the changes in these parameters were significant 1 year after CXL.

CONCLUSIONS

Although the CorVis ST device may detect changes in some biomechanical properties of cornea after the treatment of keratoconus patients with CXL, many parameters remain unchanged, and this device cannot readily be used to find the effects of CXL.

摘要

目的

使用Corvis ST设备评估圆锥角膜患者接受紫外线A/核黄素角膜交联(CXL)治疗后角膜生物力学特性的变化。

方法

本前瞻性观察性病例系列纳入了37例连续的进行性圆锥角膜患者的37只眼。使用Corvis ST在CXL治疗前、治疗后3个月和1年记录角膜生物力学参数,包括扁平角膜长度(L1和L2)、第一次和第二次扁平过程中角膜扁平时的运动速度(V1和V2)、变形幅度(DA)、角膜弯曲点之间的距离(PD)以及最高凹陷点处的角膜凹面曲率半径(R)。

结果

患者的平均年龄为23.27岁(范围19 - 31岁)。在CorVis ST角膜生物力学参数中,最高凹陷点处的L1、DA、PD和R没有显著变化。第二次扁平时的扁平角膜长度(L2)在CXL治疗后3个月有显著变化,但该参数在3个月和1年的值之间没有显著差异。角膜扁平过程中的运动速度(V1和V2)在CXL治疗后3个月没有变化,但在CXL治疗后1年这些参数有显著变化。

结论

尽管CorVis ST设备可能检测到圆锥角膜患者接受CXL治疗后角膜某些生物力学特性的变化,但许多参数保持不变,并且该设备不能轻易用于发现CXL的效果。

相似文献

1
Corneal Biomechanical Changes after Corneal Cross-Linking in Patients with Keratoconus.
J Curr Ophthalmol. 2023 Apr 29;34(4):409-413. doi: 10.4103/joco.joco_177_21. eCollection 2022 Oct-Dec.
2
Assessment of the changes in corneal biomechanical properties after collagen cross-linking in patients with keratoconus.
J Curr Ophthalmol. 2019 Mar 15;31(3):262-267. doi: 10.1016/j.joco.2019.02.002. eCollection 2019 Sep.
4
Corneal biomechanical outcome of collagen cross-linking in keratoconic patients evaluated by Corvis ST.
Eur J Ophthalmol. 2021 Jul;31(4):1577-1583. doi: 10.1177/1120672120944798. Epub 2020 Jul 25.
5
In Vivo Early Corneal Biomechanical Changes After Corneal Cross-linking in Patients With Progressive Keratoconus.
J Refract Surg. 2017 Dec 1;33(12):840-846. doi: 10.3928/1081597X-20170922-02.
7
Dynamic Scheimpflug-based assessment of keratoconus and the effects of corneal cross-linking.
J Refract Surg. 2014 Jun;30(6):408-14. doi: 10.3928/1081597X-20140513-02.
8
Keratoconus diagnosis using Corvis ST measured biomechanical parameters.
J Curr Ophthalmol. 2017 May 22;29(3):175-181. doi: 10.1016/j.joco.2017.05.002. eCollection 2017 Sep.
9
[Influence factors and differences of posterior corneal elevation measured by Pentacam system combined with Corvis ST].
Zhonghua Yan Ke Za Zhi. 2020 Feb 11;56(2):110-117. doi: 10.3760/cma.j.issn.0412-4081.2020.02.006.

引用本文的文献

本文引用的文献

2
Corneal biomechanical outcome of collagen cross-linking in keratoconic patients evaluated by Corvis ST.
Eur J Ophthalmol. 2021 Jul;31(4):1577-1583. doi: 10.1177/1120672120944798. Epub 2020 Jul 25.
3
Repeatability and comparison of new Corvis ST parameters in normal and keratoconus eyes.
Sci Rep. 2019 Oct 25;9(1):15379. doi: 10.1038/s41598-019-51502-4.
4
Topographic and Biomechanical Changes after Application of Corneal Cross-Linking in Recurrent Keratoconus.
Int J Environ Res Public Health. 2019 Oct 12;16(20):3872. doi: 10.3390/ijerph16203872.
5
Assessment of the changes in corneal biomechanical properties after collagen cross-linking in patients with keratoconus.
J Curr Ophthalmol. 2019 Mar 15;31(3):262-267. doi: 10.1016/j.joco.2019.02.002. eCollection 2019 Sep.
6
Keratoconus diagnosis using Corvis ST measured biomechanical parameters.
J Curr Ophthalmol. 2017 May 22;29(3):175-181. doi: 10.1016/j.joco.2017.05.002. eCollection 2017 Sep.
9
Dynamic Scheimpflug-based assessment of keratoconus and the effects of corneal cross-linking.
J Refract Surg. 2014 Jun;30(6):408-14. doi: 10.3928/1081597X-20140513-02.
10
Accelerated versus conventional corneal collagen crosslinking.
J Cataract Refract Surg. 2014 Jun;40(6):1013-20. doi: 10.1016/j.jcrs.2013.12.012.

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验