J Refract Surg. 2014 Jun;30(6):408-14. doi: 10.3928/1081597X-20140513-02.
To evaluate the ability of the Corvis ST (Oculus Optikgeräte GmbH, Wetzlar, Germany) to distinguish between normal and keratoconic corneas and detect the possible stiffening effect of corneal cross-linking (CXL) in vivo.
A retrospective study including 86 eyes of 86 participants (27 patients with untreated keratoconus, 27 patients with CXL-treated keratoconus, and 32 normal patients). The prospective study included 9 keratoconic eyes measured 24 days before and 97 days after CXL.
The following parameters from the Corvis ST showed a significant difference between keratoconic and normal eyes: intraocular pressure, corneal thickness, first applanation time, highest concavity (HC) radius, HC deformation amplitude, and HC deflection amplitude (analysis of variance, P < .0001). Receiver operating characteristic curve analysis showed the highest area under the curve for HC deflection amplitude (0.942, P < .0001). Prediction intervals for individual measurements within the two groups overlapped considerably. No significant difference was found between the untreated and CXL-treated keratoconus groups (analysis of variance, P > .05). Measurements performed before and after CXL in the same patients showed significant difference in HC deformation amplitude (P = .018), HC time (P = .018), and second applanation time (P = .027).
Although the corneal deformation characteristics differed between groups of keratoconic and normal eyes, this study showed that the standard parameters of the Corvis ST cannot readily be used for diagnosis of keratoconus in the individual patient or to document the effect of CXL in vivo.
评估 Corvis ST(德国 Oculus Optikgeräte GmbH 公司,威兹勒)区分正常角膜和圆锥角膜的能力,并检测角膜交联(CXL)的潜在僵硬效应。
本回顾性研究纳入 86 只眼 86 名患者(未治疗的圆锥角膜 27 例,CXL 治疗的圆锥角膜 27 例,正常患者 32 例)。前瞻性研究纳入 9 只圆锥角膜眼,于 CXL 前 24 天和后 97 天进行测量。
Corvis ST 的以下参数在圆锥角膜和正常眼中存在显著差异:眼内压、角膜厚度、首次压平时间、最大凹陷半径、最大凹陷变形幅度和最大凹陷挠度幅度(方差分析,P <.0001)。受试者工作特征曲线分析显示,最大凹陷挠度幅度的曲线下面积最高(0.942,P <.0001)。两组内各测量值的预测区间重叠较大。未治疗的圆锥角膜组和 CXL 治疗的圆锥角膜组之间无显著差异(方差分析,P >.05)。同一患者的 CXL 前后测量显示最大凹陷变形幅度(P =.018)、最大凹陷时间(P =.018)和第二次压平时间(P =.027)有显著差异。
尽管圆锥角膜组和正常组的角膜变形特征存在差异,但本研究表明,Corvis ST 的标准参数不能用于个体患者的圆锥角膜诊断,也不能用于活体评估 CXL 的效果。