IEEE Trans Ultrason Ferroelectr Freq Control. 2023 Jul;70(7):782-791. doi: 10.1109/TUFFC.2023.3280635. Epub 2023 Jun 29.
Fast imaging methods are needed to promote clinical adoption of ultrasound tomography (UST), and more widely available UST hardware could support the experimental validation of new measurement configurations. In this work, an open-source 256-element transducer ring array was developed (morganjroberts.github. io/open-UST) and manufactured using rapid prototyping, for only £2k. Novel manufacturing techniques were used, resulting in a 1.17° mean beam axis skew angle, a [Formula: see text] mean element position error, and a [Formula: see text] deviation in matching layer thickness. The nominal acoustic performance was measured using hydrophone scans and watershot data, and the 61.2 dB signal-to-noise ratio (SNR), 55.4° opening angle, 10.2 mm beamwidth, and 54% transmit-receive bandwidth (-12 dB) were found to be similar to existing systems and compatible with state-of-the-art full-waveform-inversion image reconstruction methods. The interelement variation in acoustic performance was typically < 10% without using normalization, meaning that the elements can be modeled identically during image reconstruction, removing the need for individual source definitions based on hydrophone measurements. Finally, data from a phantom experiment were successfully reconstructed. These results demonstrate that the open-UST system is accessible for users and is suitable for UST imaging research.
需要快速成像方法来促进超声断层成像(UST)的临床应用,而更广泛可用的 UST 硬件则可以支持新测量配置的实验验证。在这项工作中,开发了一种开源的 256 元换能器环形阵列(morganjroberts.github.io/open-UST),并使用快速原型制造,仅需 2000 英镑。使用了新颖的制造技术,导致平均波束轴偏转角为 1.17°,平均单元位置误差为 [Formula: see text],以及匹配层厚度偏差为 [Formula: see text]。使用水听器扫描和水波数据测量了标称的声学性能,发现 61.2dB 的信噪比(SNR)、55.4°的开口角、10.2mm 的波束宽度和 54%的收发带宽(-12dB)与现有系统相似,与最先进的全波反演图像重建方法兼容。在不使用归一化的情况下,元件间的声性能变化通常小于 10%,这意味着在图像重建过程中可以对元件进行相同的建模,无需根据水听器测量对每个源进行单独定义。最后,成功重建了来自体模实验的数据。这些结果表明,开放式 UST 系统可供用户使用,适用于 UST 成像研究。