一种新的渐进性环曲面疏水型人工晶状体的旋转稳定性,其圆柱镜度数逐渐降低。

Rotational Stability of a New Progressive Toric Hydrophobic Intraocular Lens With Gradually Decreasing Cylinder Power.

出版信息

J Refract Surg. 2023 Jun;39(6):374-380. doi: 10.3928/1081597X-20230424-02. Epub 2023 Jun 1.

Abstract

PURPOSE

To investigate the rotational stability of a new hydrophobic C-loop one-piece toric intraocular lens (IOL).

METHODS

In this retrospective multicenter case series, the Avansee Preload1P Toric Clear (Kowa Co Ltd) was implanted using a digital marking system. Its orientation was assessed with retroillumination photographs at 1 hour, 1 day, 1 week, 1 month, 3 months, and 6 months. The mean rotation at each follow-up examination and percentage of eyes with a rotation within 5° and 10° were recorded.

RESULTS

Seventy-two eyes were enrolled and completed the 3-month follow-up examination; data for the 6-month follow-up examination were obtained in 56 eyes. From the first postoperative to the 3-month examination, the mean arithmetic and absolute rotations were 0.58° ± 2.97° and 1.44° ± 2.65°, respectively. During this period, the rotation was 10° or less in 71 of 72 eyes (98.6%) and 5° or less in 67 of 72 eyes (93.1%). In the subgroup of 56 eyes with a 6-month follow-up, the mean arithmetic and absolute rotations were 0.95° ± 2.86° and 2.27° ± 1.96°, respectively, from the first to the last examination. During this period, the rotation was 10° or less in 100% of eyes and 5° or less in 53 of 56 eyes (94.6%).

CONCLUSIONS

The new toric IOL has high rotational stability. The measured values were better than the corresponding values previously reported for other toric IOLs at all times considered up to 3 months and similar to them at 6 months. It satisfies the International Organization of Standardization and American National Standards Institute requirements. .

摘要

目的

研究一种新型疏水 C 环一体式散光人工晶状体(IOL)的旋转稳定性。

方法

本回顾性多中心病例系列研究中,使用数字标记系统植入了 Avansee Preload1P Toric Clear(Kowa Co Ltd)。在术后 1 小时、1 天、1 周、1 个月、3 个月和 6 个月时,通过背面照明照片评估其方位。记录每次随访时的平均旋转角度以及旋转角度在 5°和 10°以内的眼数百分比。

结果

共纳入 72 只眼并完成了 3 个月的随访检查;56 只眼获得了 6 个月的随访检查数据。从术后第 1 天到 3 个月检查,平均算术和绝对旋转角度分别为 0.58°±2.97°和 1.44°±2.65°。在此期间,72 只眼中有 71 只(98.6%)的旋转角度为 10°或以下,67 只(93.1%)的旋转角度为 5°或以下。在 56 只具有 6 个月随访的亚组中,从第 1 次检查到最后 1 次检查,平均算术和绝对旋转角度分别为 0.95°±2.86°和 2.27°±1.96°。在此期间,所有眼的旋转角度均为 10°或以下,56 只眼中有 53 只(94.6%)的旋转角度为 5°或以下。

结论

新型散光人工晶状体具有较高的旋转稳定性。在所有考虑的时间点,测量值均优于之前报道的其他散光人工晶状体在 3 个月时的相应值,且与 6 个月时的结果相似。它符合国际标准化组织和美国国家标准协会的要求。