一种在金属过敏患者中成功植入包裹有膨体聚四氟乙烯片的植入式心脏复律除颤器的方法。

A method for successfully implanting an implantable cardioverter-defibrillator wrapped with an expanded polytetrafluoroethylene sheet in a patient with metal allergy.

作者信息

Morishita Kei, Ishihara Akiko, Unno Takatoshi, Murakami Takahiko, Okada Kensuke, Matsunaga Hiroshi, Asada Kazuo, Omori Yasutoshi, Kamoi Yoshiro, Tanaka Takahiro

机构信息

Department of Cardiology, Showa General Hospital, Tokyo, Japan.

出版信息

J Cardiol Cases. 2023 Apr 29;28(1):40-43. doi: 10.1016/j.jccase.2023.04.001. eCollection 2023 Jul.

Abstract

UNLABELLED

Contact dermatitis is a severe complication of cardiac-device implantation that may be observed in patients with metal allergies. Some studies have suggested that wrapping cardiac devices with expanded polytetrafluoroethylene (ePTFE) sheets is effective in preventing contact dermatitis. Most of these studies involved pacemakers, whereas those on implantable cardioverter-defibrillators (ICDs) are rare. Herein, we report a method for the successful implantation of an ICD wrapped with an ePTFE sheet in a patient with metal allergy. The metal part of the ICD generator was tightly wrapped with an ePTFE sheet, which was sewn with ePTFE sutures approximating the edges of the generator. After the wrapping procedure, the patient entered the operating room, and the generator and an ePTFE-coated dual-coil shock lead were implanted via a standard procedure. The shock impedance in the coil-to-can vector was high immediately after the implantation, but it reduced to less than half of its initial value over a period of two weeks post-surgery. The patient did not develop any new skin problems during the 20-month follow-up. This is a method for successfully preventing contact dermatitis; however, attention to the associated high risk of infection is required.

LEARNING OBJECTIVE

Wrapping an implantable cardioverter-defibrillator with an expanded polytetrafluoroethylene sheet was effective in preventing contact dermatitis after implantation. The shock impedance in the coil-to-can vector was high immediately after implantation but reduced to approximately half of its initial value with time.

摘要

未标注

接触性皮炎是心脏设备植入的一种严重并发症,在金属过敏患者中可能会出现。一些研究表明,用膨体聚四氟乙烯(ePTFE)片包裹心脏设备可有效预防接触性皮炎。这些研究大多涉及起搏器,而关于植入式心律转复除颤器(ICD)的研究很少。在此,我们报告一种在金属过敏患者中成功植入用ePTFE片包裹的ICD的方法。ICD发生器的金属部分用ePTFE片紧密包裹,并用ePTFE缝线缝合,使其靠近发生器边缘。包裹操作完成后,患者进入手术室,通过标准程序植入发生器和ePTFE涂层双线圈除颤导线。植入后立即,线圈至罐体向量的电击阻抗较高,但在术后两周内降至初始值的一半以下。在20个月的随访期间,患者未出现任何新的皮肤问题。这是一种成功预防接触性皮炎的方法;然而,需要注意相关的高感染风险。

学习目标

用膨体聚四氟乙烯片包裹植入式心律转复除颤器可有效预防植入后接触性皮炎。植入后立即,线圈至罐体向量的电击阻抗较高,但随时间降至初始值的约一半。